Guidance and manufacture's declaration – electromagnetic immunity –
Guidance and manufacture's declaration – electromagnetic immunity
The ECG100G ECG is intended for use in the electromagnetic environment specified below. The customer or the user of
ECG100G ECG should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60Hz)
magnetic field
IEC61000-4-8
NOTE
U
is the a.c. mains voltage prior to application of the test level.
T
for all EQUIPMENT and SYSTEMS
IEC 60601 test level
6 kV contact
8 kV air
2 kV for power supply
lines
1 kV for signal lines
1 kV differential mode
2 kV common mode
<5% U
T
(>95% dip in U
)
T
for 0.5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
<5% U
T
(>95% dip in U
)
T
for 5 sec
3A/m
User Manual
Compliance level
6 kV contact
8 kV air
2 kV for power supply
lines
1 kV for signal lines
1 kV differential mode
2 kV common mode
<5% U
T
(>95% dip in U
)
T
for 0.5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
<5% U
T
(>95% dip in U
)
T
for 5 sec
3A/m
25
Electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tile. If floor
are covered with synthetic
material, the relative humidity
should be at least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the
user of the ECG100G ECG
requires continued operation
during power mains dip &
interruptions, it is
recommended that the
ECG100G ECG be powered
from an uninterruptible power
supply or a battery.
Power frequency magnetic
fields Should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.