Calibration Without I.r. Radiometer; Maintenance - UGO BASILE 37370 Instruction Manual

Plantar test hargreaves apparatus
Table of Contents

Advertisement

Depress the F1 (UP ) or F2 (DWN) key to reach the value indicated on the Radiometer,
21mW/cm
, then depress F3 (STP) to save the calibration and confirm it by the F4 (OK )
2
key.
Now depress F3 (MED) key, then F3
(STR).
Depress the F1 (UP ) or F2 (DWN) key to
reach the value indicated on the Radiome-
ter, 190mW/cm
, then save and confirm it
2
by the F3 (STP) and F4 (OK ) keys.
9.3

Calibration without I.R. Radiometer

Calibrating the Plantar Test without the I.R. Radiometer, means to recall the default cali-
bration value set at our factory during the testing phase.
From the Calibration Menu, depress the F1 (DEF) key and proceed by the F3 (YES) key,
to confirm the change.
The default calibration value is thus recalled.
10

MAINTENANCE

While any service of the instrument ought to be carried out by Ugo Basile personnel or by
qualified personnel authorized by UGO BASILE organization, this section of the instruc-
tion manuals describes normal maintenance procedures which can be carried out at your
facility.
37370 Instruction Manual (Rev. 0)
Page 23