Calibration Without I.r. Radiometer; Maintenance; Electrical - UGO BASILE 37360 Instruction Manual

Tail-flick unit
Table of Contents

Advertisement

Depress the F1 (UP ) or F2 (DWN) key to
reach the value indicated on the Radiome-
2
ter, 260mW/cm
, then save and confirm it
by the F3 (STP) and F4 (OK ) keys.
9.3

Calibration without I.R. Radiometer

Calibrating the Tail Flick Unit without the I.R. Radiometer, means to recall the default cal-
ibration value set at our factory during the testing phase.
From the Calibration Menu, depress the F1 (DEF) key and proceed by the F3 (YES) key,
to confirm the change.
The default calibration value is thus recalled.
10

MAINTENANCE

While any service of the instrument ought to be carried out by Ugo Basile personnel or by
qualified personnel authorized by UGO BASILE organization, this section of the instruc-
tion manuals describes normal maintenance procedures which can be carried out at your
facility.
UNPLUG THE MAINS CORD BEFORE CARRYING OUT
ANY MAINTENANCE JOB!
10.1

Electrical

To inspect and/or replace the fuses, disconnect the mains cable first! Insert a miniature
screwdriver in the slot indentation, see paragraph 4.5.1, and snap out the slide which
houses the fuses.
The Tail Flick Unit universal mains input accepts any voltage from 85 to 264 Volt AC., for
which we recommend the use of 2A fast-blow fuses (type F2A).
Snap in the fuse slide: the mechanical "click" ensures that it is locked.
37360 Instruction Manual (Rev. 0)
260 mW/cm sq
Page 22