Write Commands "Cvc 3000 - vacuubrand ME 4C NT VARIO Instructions For Use Manual

Speed controlled chemistry diaphragm pumps and chemistry pumping units
Hide thumbs Also See for ME 4C NT VARIO:
Table of Contents

Advertisement

Write commands "CVC 3000"

Operation
Command
function
OUT_MODE
Attention: If control is running
only switching from 1 to 2, 2 to 3
and 3 to 2 is possible with takeo-
ver of the set vacuum.
configuration
OUT_CFG
(bus monitoring)
set vacuum
OUT_SP_1
set vacuum
OUT_SP_V
with venting
speed
OUT_SP_2
start-up pressure
OUT_SP_3
delay
OUT_SP_4
switch off pressure
OUT_SP_5
switch-off time
OUT_SP_6
open program
OUT_SP_PL
store program
OUT_SP_PS
time
OUT_SP_P1y
pressure
OUT_SP_P2y
valve
OUT_SP_P3y
Step
OUT_SP_P4y
Auto
OUT_SP_P5y
Documents are only to be used and distributed completely and unchanged. It is strictly the users' responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999164 / 19/05/2009
page 48 of 75
Parameter
Description
X
0: VACUULAN
1: Pump down
2: Vac control
3: Auto mode
4: Program
XXXX
0: acoustic signal off
1: acoustic signal on
0: Autostart off
1: Autostart on
0: pressure unit mbar
1: pressure unit Torr
2: pressure unit hPa
0.....C: language*
XXXX
unit according to preselection
(0001 to 1060 mbar (hPa) or 0001 to 0795 Torr)
XXXX
unit according to preselection
(0001 to 1060 mbar (hPa) or 0001 to 0795 Torr)
XXX
speed in %, (100% = "HI")
XXXX
unit according to preselection
(0001 to 1060 mbar (hPa) or 0001 to 0795 Torr)
XX:XX
hh:mm (hours:minutes)
XXXX
unit according to preselection
(0001 to 1060 mbar (hPa) or 0001 to 0795 Torr)
XX:XX
hh:mm (hours:minutes)
X
programm 0.....9
X
programm 0.....9
XX:XX:XX
total runtime until program step y (0.....9) or
+XX:XX:XX
time for program step y (0.....9) (additive)
XXXX
pressure at program step y (0.....9)
X
0: no venting valve in program step y (0.....9)
X
1: venting valve in program step y (0.....9)
0: no "Step" in program step y (0.....9)
X
1: "Step" in program step y (0.....9)
0: no "Auto" in program step y (0.....9)
1: "Auto " in program step y (0.....9)
2: "Auto " in program step y (0.....9)
30: sensitivity: low
31: sensitivity: normal
32: sensitivity: high
(optional)
* Language:
0: German
1: English
2: French
3: Italian
4: Spanish
5: Turkish
6: Korean
7: Chinese
8: Portuguese
9: Russian
A: Polish
B: Dutch
C: Japanese

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents