Write Commands "Cvc 2000 - vacuubrand ME 4C NT VARIO Instructions For Use Manual

Speed controlled chemistry diaphragm pumps and chemistry pumping units
Hide thumbs Also See for ME 4C NT VARIO:
Table of Contents

Advertisement

Write commands "CVC 2000"

Operation
function
set vacuum
set vacuum
with venting*
selected frequency
pressure for restart
(VACUU•LAN)
delay
vacuum for automatic
switching off
time for automatic
switching off (VACUU•LAN)
starting process control START
stopping process control STOP
Remote operation**
driving venting valve
* Pressure setting with venting is only possible in function "Vacuum control" if an venting valve is
connected and configurated and vacuum control is started. The venting valve opens automatically
if the actual pressure is at least 10 mbar below the preset pressure. Automatic venting becomes
inactive if vacuum control is stopped (STOP or VENT), setting a pressure value using the command
OUT_SP_1 or if the function is changed. Activate the command OUT_SP_V again if necessary.
** If remote operation is selected or deselected, the user has to ensure that no dangerous status of
the system due to the change of the mode of operation can occur and to provide appropriate safety
measures, especially if selecting remote operation interferes with a locally operated active process.
Connection plug arrangement
Documents are only to be used and distributed completely and unchanged. It is strictly the users' responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999164 / 19/05/2009
page 44 of 75
Command
Parameter
OUT_MODE
X
OUT_SP_1
XXXX
OUT_SP_V
XXXX
OUT_SP_2
XX.X
OUT_SP_3
XXXX
OUT_SP_4
XX:XX
OUT_SP_5
XXXX
OUT_SP_6
XX:XX
X
REMOTE
X
OUT_VENT
X
2: RxD
3: TxD
4: DTR
Description
1: continuous pumping
2: vacuum control without automatic
3: vacuum control with automatic
30: sensitivity: low
31: sensitivity: normal
(optional)
32: sensitivity: high
unit according to preselection
(0001 to 1060 mbar (hPa) or 0001 to 0795 Torr)
unit according to preselection
(0001 to 1060 mbar (hPa) or 0001 to 0795 Torr)
motor frequency in Hz (01.0 to 60.0 in steps
of 0.5 Hz or 99.9 for "HI")
unit according to preselection
(see OUT_SP_1)
hh:mm (hours:minutes)
unit according to preselection
(0001 to 1060 mbar (hPa) or 0001 to 0795 Torr)
hh:mm (hours:minutes)
1: termination of process control
2: termination of process control and storage of
the current pressure as new set point
0: set controller to local operation
1: set controller to remote operation
0: close venting valve (not automatically)
1: open venting valve
(process control stopped)
5: Mass
7: RTS
8: CTS
9: +5V (Bluetooth)