Chlorine, Total - Hanna Instruments HI83303 Instruction Manual

Aquaculture photometer
Hide thumbs Also See for HI83303:
Table of Contents

Advertisement

8.9. CHLORINE, TOTAL

SPECIFICATIONS
Range
Resolution
Accuracy
Light Source
Method
REQUIRED REAGENTS
POWDER:
Code
HI93711‑0
LIQUID:
Code
HI93701A‑T
HI93701B‑T
HI93701C‑T
REAGENT SETS
HI93701‑T
HI93711‑01
HI93711‑03
For other accessories see page 87.
MEASUREMENT PROCEDURE
• Select the
Chlorine (Total)
the Method Selection section (see page 17).
• Fill the cuvette with 10 mL of unreacted sample (up to the mark)
and replace the cap.
• Place the cuvette into the holder and close the lid.
52
0.00 to 5.00 mg/L (as Cl
2
0.01 mg/L
±0.03 mg/L ±3% of reading at 25 °C
LED with narrow band interference filter @ 525 nm
Adaptation of the EPA DPD method 330.5.
Description
Total Chlorine Reagent
Description
Total Chlorine Reagent A
Total Chlorine Reagent B
Total Chlorine Reagent C
Reagents for 300 tests (liquid)
Reagents for 100 total tests (powder)
Reagents for 300 total tests (powder)
method using the procedure described in
)
Quantity
1 packet
Quantity
3 drops
3 drops
1 drop
10 mL

Hide quick links: