Thin trms, bluetooth, autoranging multimeter

Advertisement

IMPORTANT: RECEIVING INSTRUCTIONS
Visually inspect all components for shipping damage. If you find damage, notify the carrier at once.
Shipping damage is NOT covered by warranty. The carrier is responsible for all repair or replacement costs resulting from damage in shipment.
Read this owners manual thoroughly before use and save.
Figure 1
1
2
3
A
400A
A
400A
4
Rotary Switch
OFF
Power off
AC/DC Voltage Measurement
Diode / Continuity / Resistance
Measurement
Hz%
Frequency / Duty Cycle
A
400A
A
400A
Figure 2
9
8
7
5
6
Capacitance
A
AC/DC Current Measurement
400A
NCV
Non-contact Voltage Measurement
A
400A
˚C ˚F
°C/°F Temperature Measurement
OPERATING INSTRUCTIONS
Thin TRMS, Bluetooth, Autoranging Multimeter
1.0 METER FUNCTIONS & DISPLAY SYMBOLS
1.1 FUNCTIONS (Figure 1)
1. Display: Backlit 3999 Count LCD screen
2. SELECT Button: Press to switch between:
• DC voltage and AC voltage measurement functions
• Resistance, diode, and continuity test functions
• Frequency and Duty Cycle measurement functions
• DC current and AC current measurement functions
• Celsius temperature and Fahrenheit temperature measurement functions
Hold for about 2 seconds to enable or disable Bluetooth communication
3. RELATIVE Button: Press to enter or exit Relative mode. Hold for about
2 seconds to start or stop recording readings in the memory of the meter.
4. 10A Terminal: Red test lead input jack for current measurements.
5. COM Terminal: Black test lead input jack common to all measurements.
6. INPUT Terminal: Red test lead input jack for all measurements
except current measurements.
7. Rotary Switch: Turn to power the meter on and off and select the desired
measurement function. To save battery power, turn to the "OFF" position when
the meter is not in use. See "ROTARY SWITCH" section for function descriptions.
8. MAX/MIN Button: Press to enter or exit the Min/Max recording mode.
9. HOLD Button: Press to enter or exit the Data Hold mode.
Hold for about 2 seconds to turn the backlight on or off. .
1.1 METER FUNCTIONS
Meter type:
Auto
Display Count:
3999
Sampling Rate:
About 2 to 3 times/sec
Battery:
Requires 2 AAA batteries
Input impedance:
10 Meg Ohm
AC Volt Ranges:
4V, 40V, 400V, 600V, best accuracy (1.0%+5)
DC Volt Ranges:
400mV, 4V, 40V, 400V, 600V, best accuracy (0.5%+5)
AC Amps:
4A, 10A, best accuracy (1.0%+5)
DC Amps:
4A, 10A, best accuracy (1.0%+5)
Resistance Ranges:
400ohm, 4kohm, 40kohm, 400kohm, 4M ohm,
and 40M ohm, best accuracy (0.8%+3)
Over Range Indication:
Displayed value > 3999 or the input
measurement range, displays OL
Frequency Ranges:
9.999Hz, 99.99Hz, 999.9Hz, 9.999kHz, 99.99kHz,
999.9kHz, 9.999MHz, best accuracy (1.0%+5)
Duty Cycle Range:
5%-95%
Capacitance Ranges:
9.999nF, 99.99nF, 999.9nF, 9.999uF, 99.99uF, 999.9uF,
best accuracy (2.5%+5)
Temperature Ranges:
-4°F–1832°F (-20°- 1000°C)
Diode Test:
The approximate forward voltage drop of the
diode will be displayed
Continuity Test:
Buzzer will sound if the resistance is < ~30Ω
Non-contact Volt Ranges:
50 VAC, 120 VAC, 500 VAC
Polarity Indication:
"-" is displayed for negative polarity
Overload Protection:
10A/600V Fast fuse
Agency Approvals:
ETL, CE (IEC/EN61010:, CAT III 600V, Pollution Degree 2)
Operating Temperature:
32°F–104°F (0°C–40°C)
Relative Humidity:
< 75%
Storage Temperature:
14°F–122°F (-10°C–50°C)
IP Degree:
IP20
Dimension:
176mm x 81mm x 17mm
Weight:
Around 180g (without batteries)
Altitude:
Maximum 2000m
Warranty Info:
Limited lifetime warranty
1
DM6850T

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the DM6850T and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Sperry instruments DM6850T

  • Page 1 OPERATING INSTRUCTIONS Thin TRMS, Bluetooth, Autoranging Multimeter DM6850T IMPORTANT: RECEIVING INSTRUCTIONS Visually inspect all components for shipping damage. If you find damage, notify the carrier at once. Shipping damage is NOT covered by warranty. The carrier is responsible for all repair or replacement costs resulting from damage in shipment.
  • Page 2: Safety Warnings

    1.2 DISPLAY SYMBOLS (Figure 2) Autorange MAX MIN 11. NCV: Non-contact : Continuity test is selected 1 . 8 . 8 . 8 AC Voltage detection is selected : Diode test is selected Autorange MAX MIN : The battery is low and must 1 .
  • Page 3 WARNING • Never attempt to make any measurement if any abnormal conditions are noted, such as broken case, cracked test leads and exposed metal part. • Do not turn the function selector switch with plugged in test leads connected to the circuit under test. •...
  • Page 4: Specification

    3. SPECIFICATION 3.1 Measuring range & accuracy (Accuracy guaranteed at 18°C - 28°C, humidity <75%) AC CURRENT FREQUENCY RANGE RESOLUTION ACCURACY RANGE RESOLUTION ACCURACY 0.001A ±(1.5% + 5) 99.99Hz 0.01Hz 0.01A ±(3.0% + 10) 999.9Hz 0.1Hz ±(10% + 5) - Frequency Response: 50Hz - 400Hz 9.999kHz 0.001kHz 99.99kHz...
  • Page 5: Backlit Display

    4.0 FUNCTIONS 4.1 MAX/MIN BUTTON 1. Press MAX/MIN once to enter MAX mode. The symbol “MAX” will appear on the display and the display will show the maximum reading that the meter measures while in this mode. 2. Press MAX/MIN again to enter MIN mode. The symbol “MIN” will appear on the display and the display will show the minimum reading that the meter measures while in this mode.
  • Page 6 DC Volts To avoid personal injury or damage to the Meter, do not attempt to measure voltages higher than 600 VDC. 1. Insert the black (negative) test lead into the COM input terminal. 2. Insert the red (positive) test lead into the INPUT terminal to the right of the COM terminal. 3.
  • Page 7: Frequency/Duty Cycle

    5.3 RESISTANCE/CONTINUITY/DIODE Resistance When measuring resistance always make sure the power to the circuit is off. 1. Insert the black (negative) test lead into the COM input terminal. 2. Insert the red (positive) test lead into the INPUT terminal to the right of the COM terminal. 3.
  • Page 8 Figure 6 bar to toggle between meter selections 6.0 SMART DEVICE & APPLICATION GPS location 1. The Sperry DM6850T has the ability to be connected Datalogging symbol and seconds to your smart device via a Bluetooth connection. active displayed when datalogging 2.
  • Page 9: Download The Application

    1. Locate the Sperry SMARTmeter app. 2. Touch the icon to open. 6.3. CONNECTING YOUR SMART DEVICE After the opening the To toggle the meter app, select DM6850T selected when already in the app, tap the blue bar at the top Press the blue flashing It may be necessary “CONN”...
  • Page 10: Data Logging

    2. Open the SMARTmeter app on your smart device. 3. Tap the blue banner at the top of the screen on the App to change the meter you have selected. Switch to DM6850T 4. Touch the “CONN” button on the bottom right of the screen on your smart device and follow the instructions on the screen.
  • Page 11: Data File Management

    6.6 DATA FILE MANAGEMENT • Open measurements previously saved to your phone/tablet. • Save currently loaded measurements to your phone/tablet. 6.6.1 Rename & Save 1. Once data has been collected it can be given a unique file name by pressing the “File” button. 2.
  • Page 12 6.6.8 Remarks 3. Enter the comment and press “Back” 1. To add a comment about an individual data point to return to the previous screen. press the blank space in the remark column. 4. Your comment will now be displayed 2.
  • Page 13: Battery Replacement

    (see section 6.0 SMART DEVICE & APPLICATIONS for instructions). 6. In the SmartMeter app, go to the main screen. Select Get Local. Datalogged data stored on the DM6850T will be displayed in the app. This data can be managed and/or shared (See Section 6.6 Data File Management and Section 6.7 Data File Sharing above).
  • Page 14 Test leads, fuses, batteries and calibration are not covered under any implied warranty. “Lifetime” of products that are no longer offered by Sperry will be either repaired or replaced with an item of Sperry Instruments choice of similar value. Lifetime is defined as 5 years after Sperry discontinued manufacturing the product, but the warranty period shall be at least ten years from date of purchase.

Table of Contents