1.2 Specifications

SPECIFICATION / MODEL
Sensitivity
Accuracy
Vernier vertical sensitivity
Frequency bandwidth
Rise time
Input impedance
Square wave characteristics
DC balance shift
Linearity
VERTICAL
AXIS
Vertical modes
Chopping repetition frequency
Input coupling
Maximum input voltage
Common mode rejection ratio
Isolation between channels
(At 5 mV/DIV range)
CH1 signal output
CH2 INV BAL.
Dynamic range
Signal delay
Triggering source
TRIGGERING
Coupling
Polarity
Page 2
OD-514B, OD-515B & OD-545B INSTRUCTION MANUAL
OD-514B
1 mV ∼ 5V/DIV, 12 steps in 1-2-5 sequence
5 mV ∼ 5V/DIV: ≤ 3%, 1 mV ∼ 2 mV/DIV: ≤ 5% (10ºC to 35ºC)
(5 DIV at the centre of display)
To 1/2.5 or less of panel-indicated value.
5mV~2V/DIV:
DC~40MHz (-4 dB)
5V/Div:DC~40MHz (-5dB)
1mV~2mV/DIV:
DC~10MHz (-3dB)
AC coupling: Low limit frequency 10Hz.
(With reference to 100kHz, 8 DIV. Frequency response with –3 dB.)
5 mV ∼ 5 V/DIV: ≈ 8.75 ns
1 mV ∼ 2 mV/DIV: ≈ 35 ns
1 M
Overshoot: ≤ 7% (At 10 mV/DIV range)
(5 DIV at the centre of display)
Other distortions and other ranges: 5% added to the above value. (10ºC to 35 ºC)
5 mV ∼ 5 V/DIV: ±0.5 DIV, 1 mV ∼ 2 mV/DIV: ±2.0 DIV
< ±0.1 DIV of amplitude change when waveform of 2 DIV at graticule centre is
CH1:
CH1 single channel.
CH2:
CH2 single channel.
DUAL: CH1 & CH2 channels. CHOP/ALT are auto-set by TIME/DIV switch.
(CHOP:0.5 s ∼ 5 ms/DIV, ALT: 2 ms ∼ 0.1
When CHOP switch is pushed in, the two traces are displayed in the
CHOP mode at all range.
ADD: CH1 + CH2 algebraic addition.
400 V (DC + AC peak), AC: frequency 1 kHz or lower.
50:1 or better at 50 kHz sinusoidal wave.
(When sensitivities of CH1 and CH2 are set equally)
>
1000:1 at 50 kHz
>30:1 at 35 MHz
Approx. 100 mV/DIV without termination, 50 mV/DIV with 50
Bandwidth (-3 dB): OD-514B: 20 MHz, OD-515B & OD-545B: 40 MHz
Balanced point variation: ≤ 1 DIV (Reference at centre graticule.)
> 8 DIV at 20 MHz
> 4 DIV at 40 MHz
---
CH1, CH2, LINE, EXT (CH1 and CH2 can be selected only when the vertical
mode is DUAL or ADD). In ALT mode, if the TRIG. ALT switch is pushed in, it can
be use for alternate triggering of two different source. )
(TV-V/TV-H can be auto-set by TIME/DIV range. TV-V: 0.5 s-0.1 ms/DIV;
OD-515B
5 mV ∼ 5 V/DIV: DC ∼ 60 MHz,
1 mV ∼ 2 mV/DIV: DC ∼ 15 MHz
5 mV ∼ 5 V/DIV:≈ 5.8 ns,
1 mV ∼ 2 mV/DIV: ≈ 23 ns
:
±2% // Approx. 25 pF
moved vertically.
P
s/DIV)
Approx. 250 kHz
AC, DC, GND
> 1000:1 at 50 kHz
> 30:1 at 60 MHz
> 8 DIV at 50 MHz
> 5 DIV at 60 MHz
---
AC, HF-REJ, TV, DC
P
P
s ∼ 0.1
TV-H: 50
s/DIV)
+ / −