Operation Method; Single-Channel Operation - Promax OD-514B Instruction Manual

Dual-channel oscilloscopes
Table of Contents

Advertisement

Remove the probe from CAL 2 Vpp terminal [1]. Set switch VERTICAL-MODE [39]
to CH2 position and repeat the process for CH2, each one with its own probe.
Now the oscilloscope is ready to be used.

4.3 OPERATION METHOD

4.3.1 Single-Channel Operation

Single trace and base time operation and internal triggering is the most basic
operating mode of the OD-514B, OD-515B & OD-545B. Select this mode when there is
only one signal to be observed, without the inconvenience of other traces on the CRT
screen. Due it is a two channel oscilloscope, any of these two channels can be selected.
CH1 is provided with and output terminal: use the CH1 output terminal on the rear panel to
measure signal frequency using a frequency counter while observing waveform. CH2 have
a switch that inverts polarity, although it adds flexibility, it has no use when operating in
single trace mode.
OD-514B, OD-515B and OD-545B are set for single channel operation as follows:
1. Set the following controls of the instrument as shown below. Observe the
selected triggering source (CH1 or CH2 SOURCE) corresponds to the selected
channel (CH1 or CH2 V MODE).
VERT MODE
   
TRIGGER MODE
SWP. UNCAL
   
2. Position trace near the centre of the screen by means of the
control [37] or
3. Connect the signal to be observed to the appropriate input terminal CH1 [12] or
CH2 [16] and adjust the appropriate VOLTS/DIV selector [10] or [14] so that
signal waveform is completely displayed on the screen.
WARNING
Do not apply signal higher than 400 V (DC + AC peak)
Page 20
CONTROL
I/O
[9]
[39]
[40][37]
POSITION
[10][14]
VOLTS/DIV
VARIABLE
[13][17]
AC/GND/DC
[11][15]
SOURCE
[26]
[28]
TIME/DIV
[18]
[30]
[34]
POSITION
X-Y
[27]
   
POSITION control [40] (depending on the channel selected).
OD-514B, OD-515B & OD-545B INSTRUCTION MANUAL
STATE
ON (engage position)
CH1 (CH2)
Mid-position
0.5 V/DIV
CAL clockwise position
AC
CH1 (CH2)
AUTO
0.5 ms/DIV
Released
Mid-position
Released
Table 2.-
   
POSITION
November 2000