7.6 Lubricating the Chair

Disinfect and clean the chair before applying lubricant.
There are two identical lubrication points on the chair. They are located
where the brass bushing in the chair lock bar slot meets the stainless
steel slide in the chair leg (Figure 31).
Lubricate each lubrication point with one drop of SAE 30-weight oil,
then move the lock bar back and forth a few times to work the oil into
the slide. Remove any excess lubricant with SD-20 Cleaner.

Note:
To order SD-20™ Cleaner in North America, contact the Spartan
Chemical Company, Maumee, Ohio, USA at 1.800.537.8990 or visit
www.spartanchemical.com
on the internet.
Lubricating the track system can result in injury to patient
and/or operators. Never lubricate the tracks or belts.
Lubricating parts that should not be lubricated allows dirt
and foreign particles to collect on those parts, resulting in
damage. Lubricate only the reference points shown.

7.7 Do Not Lubricate Track System

Never lubricate the tracks, belts, or any other part of the track
system. Lubricants on the tracks or belts can cause the chair to perform
unpredictably, resulting in injury to the patient and/or operators.
A label (Figure 32) instructing the user not to lubricate the track system
is affixed to each track. If these labels become damaged or worn, replace
them immediately. See "Parts and Service" on page 35.
If track or belts pick up oil or grease during use or while the chair is being
serviced, clean the track and belts with SD-20 All-Purpose Cleaner before
using the chair again. See "Cleaning Tracks and Belts" on page 27.
© Ferno-Washington, Inc. / 234-3298-07 / July 2018