Brief Introduction - ETCR ETCR2000+ Manual

Earth resistance tester
Table of Contents

Advertisement

This instrument was designed by technical department of
ETCR Electronic Technology Company. Its performance is mainly
reflected in:
u Breakthrough in self-test the boot a long time to wait, start
immediately into the test.
u Breakthrough relay self-test mode, using the most
advanced processing algorithms and digital integration
technology, a fully intelligent.
u Break the old product to heavy issues, more in line with
characteristics of handheld devices.
u New design, panel operation with 6 buttons, better
performance.
u An increase of sound and light alarm, "beep—beep--beep
--" alarm sound.
u Increase the interference signal recognition indicator.
u Improved anti-jamming capability and test stability.
u Stored data: 99 Units.
u Wider range: 0.01Ω-1200Ω
u Lower power consumption: Maximum operating current
not exceeding 50mA.
ETCR series of Pincers Earth Tester is widely used in the
grounding
telecommunications, meteorology, oilfield, construction and the
II. Brief Introduction
resistance
measurement
-3-
of
the
power,

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ETCR2000+ and is the answer not in the manual?

Questions and answers

Luis A. Flores Torres
April 15, 2025

CALIBRATION TIME

1 comments:
Mr. Anderson
April 15, 2025

The calibration for the ETCR2000+ occurs automatically during boot-up. The device performs auto-calibration after pressing the POWER key and shows "OL Ω" upon successful completion. The exact duration of the calibration time is not specified.

This answer is automatically generated

Ado Dahiru
April 20, 2025

Unit does not switch on,batteries are good.

1 comments:
Mr. Anderson
May 10, 2025

If the ETCR2000+ unit does not switch on despite having good batteries, possible issues include:

- Incorrect battery polarity: Ensure batteries are installed correctly.
- Poor contact of battery contacts: Check and clean or replace the battery contacts.
- Wrong battery type: Use the correct battery type.
- A break in the battery harness: Perform a continuity test; replace the harness if no continuity.
- Displaced POWER button: Re-assemble the button.
- Poor contact of power plug: Re-plug or replace the plug.
- Defective PCB or circuit component: Replace the PCB if current consumption is about 100mA or more at 6V.