2. Description of Special Symbols

Note: "*" is limited to C+.
.
Symbol of an open jaw: As a jaw is in the open state, the
symbol shows. At this point, trigger may be artificially pressed,
or the jaws have been seriously polluted, and can no longer
continue to measure.
⑵ "Er" Boot error symbol, May be pressing trigger when boot or
.
jaw has been opened.
.
Symbol of low battery & voltage: when the battery
voltage is lower than 5.3V, the symbol shows. At this time, it
cannot guarantee accuracy of the measurements. Batteries
should be replaced.
⑷ "OL Ω" symbol indicates that the measured resistance has
.
exceeded the upper limit of the Meter.
⑸ "L0.01Ω" symbol indicates that the measured resistance has
.
exceeded the lower limit of the Meter.
* .
⑹ "OL A" symbol indicates that the measured current has
exceeded the upper limit of the Meter.
.
Alarm symbol: when the measured value is greater than
the critical value of alarm setting, the symbol flashes, and the
meter issued by intermittent "beep--beep –beep--" sound.
⑻ MEM Symbol of full data storage: memory is full of data units
.
of 99, and can no longer continue to store data. MEM symbol
flashes.
⑼ MR Symbol of access to data: to display in an access to data,
.
also including the number of data.
-9-

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ETCR2000+ and is the answer not in the manual?

Questions and answers

Luis A. Flores Torres
April 15, 2025

CALIBRATION TIME

1 comments:
Mr. Anderson
April 15, 2025

The calibration for the ETCR2000+ occurs automatically during boot-up. The device performs auto-calibration after pressing the POWER key and shows "OL Ω" upon successful completion. The exact duration of the calibration time is not specified.

This answer is automatically generated

Ado Dahiru
April 20, 2025

Unit does not switch on,batteries are good.

1 comments:
Mr. Anderson
May 10, 2025

If the ETCR2000+ unit does not switch on despite having good batteries, possible issues include:

- Incorrect battery polarity: Ensure batteries are installed correctly.
- Poor contact of battery contacts: Check and clean or replace the battery contacts.
- Wrong battery type: Use the correct battery type.
- A break in the battery harness: Perform a continuity test; replace the harness if no continuity.
- Displaced POWER button: Re-assemble the button.
- Poor contact of power plug: Re-plug or replace the plug.
- Defective PCB or circuit component: Replace the PCB if current consumption is about 100mA or more at 6V.

This answer is automatically generated