Venous Blood
96.4% of TRUE METRIX AIR venous values performed by healthcare professionals (HCP) fell
within 0.83 mmol/L of the YSI results at glucose levels < 5.55 mmol/L and within 15% at glucose
levels > 5.55 mmol/L.
Venous Samples (HCP vs. YSI) for glucose concentrations < 5.55 mmol/L
Within
+ 0.28 mmol/L
16/50 (32%)
Venous Samples (HCP vs. YSI) for glucose concentrations > 5.55 mmol/L
Within + 5%
33/174 (19%)
Venous Samples for glucose concentrations between 1.1-33.3 mmol/L
Parkes Error Grid: 100% of individual venous glucose measured values performed by healthcare
professionals fell within Zone A of the Parkes Error Grid (PEG).
For Consumers
99% of TRUE METRIX AIR fingertip values performed by users fell within 0.83 mmol/L of the YSI
results at glucose levels < 5.55 mmol/L and within 15% at glucose levels > 5.55 mmol/L.
Fingertip Samples (User vs. YSI) for glucose concentrations < 5.55 mmol/L
Within
+ 0.28 mmol/L
9/18 (50%)
Fingertip Samples (User vs. YSI) for glucose concentrations > 5.55 mmol/L
Within + 5%
39/82 (47.6%)
Fingertip Samples for glucose concentrations between 1.1-33.3 mmol/L
Parkes Error Grid: 100% of individual fingertip glucose measured values performed by users fell
within Zone A of the Parkes Error Grid (PEG).
Within
+ 0.56 mmol/L
39/50 (78%)
Within + 10%
100/174 (57.5%)
Within + 0.83 mmol/L and + 15%
216/224 (96.4%)
Within
+ 0.56 mmol/L
17/18 (94.4%)
Within + 10%
65/82 (79.3%)
Within + 0.83 mmol/L and + 15%
99/100 (99%)
57
Within
+ 0.83 mmol/L
50/50 (100%)
Within + 15%
166/174 (95.4%)
Within
+ 0.83 mmol/L
18/18 (100%)
Within + 15%
81/82 (98.8%)