9. With test strip still in meter, touch
Sample Tip of test strip to top of
drop of control solution. Allow
drop to be drawn into test strip.
Remove test strip from drop when
meter beeps and begins testing.
10. Dashes appear across the Display to
show meter is testing.
Note: If meter does not beep and begin testing soon
after drawing up sample, release and discard
test strip. Repeat test with new test strip. If
problem persists, see Troubleshooting.
11. Compare meter
result to Control
Test range printed
on test strip vial
label for level of
control solution you
are using. If result is
in range, system can
be used for testing
blood. If result does
not fall within range,
repeat test using a
new test strip.
Note: Control Test result shows the Control Symbol in the
Display.
Control Solution
Bottle Label
Control Symbol (Example only. Does not
represent actual Control Test ranges)
20
Meter Testing
Test Strip
Vial Label