How to Test Control Solution

Use ONLY TRUE METRIX Control Solution with
the TRUE METRIX AIR Meter and TRUE METRIX
Test Strips.
1. Check dates on control solution label
and test strip vial label. Do not use
control solution or test strips if either
Use By Dates have passed. (control
solution - 3 months after first
opening or date next to
label; test strips - after open vial Use
By date (see test strip Instructions
for Use) or date next to on vial
label.) Discard expired products and
use new products.
2. Allow control solution, vial of test strips and
meter to adjust to room temperature. Write date
first opened on both control solution bottle label
and test strip vial label when using for the first
time.
3. Gently swirl or invert control solution bottle to
mix.
DO NOT SHAKE!
4. Remove one test strip from vial. Close test strip
vial immediately. Use test strip quickly after
removal from vial.
on bottle