How To Care For This Product - ADC Adtemp 427 Temple Touch Use, Care & Maintenance

Table of Contents

Advertisement

Er.2: If the temperature did not register because of poor contact,
patient movement or poor placement, the device will beep 5 times
and display "Er.2"
Remedy: Turn the unit OFF, wait 1 minute, turn it ON again and retake tempera-
ture using proper technique.
Er.3: If the unit malfunctioned, the device will beep 5 times and
display "Er.3"
Remedy: Do not use the thermometer – Call 1-800-232-2670 for
service.
Low Battery: Battery icon is on continuously.
Remedy: The battery has enough remaining life for approxi-
mately 10 more measurements
Drained Battery: Battery icon appears as unit is turned on and
the device turns itself OFF.
Remedy: Replace battery.
Standards:
Complies with EN 12470-3 and ASTEM E 1112-00

How to care for this product:

• Wipe the tip and probe with a cotton swab moistened with 70% isopropyl alco-
hol before & after each use. After cleaning, wait at least 2 minutes before taking a
measurement for the unit to re-adjust to room temperature.
• Use only a soft dry cloth to clean the body of the thermometer (if necessary).
Never use an abrasive cleanser or submerge the thermometer in water or other liq-
uids.
6