Philips Trilogy Evo Universal Instructions For Use Manual page 69

Table of Contents

Advertisement

4.
In the Data Transfer window, tap the procedure you would like to perform and then follow the prompts on the screen:
Function
Export Data - USB
Export Data - Bluetooth
Export Alarm & Event Log
Install Prescriptions
Install Software Update
As data is transferred, a data transfer icon appears in the Status Bar.
USB Data Transfer Icon
Description
Export details about the patient's therapy to a storage device connected to the USB
port. Data is identified by the device serial number.
Data type
Changes to the ventilator, alarms, prescriptions, accessories
and oxygen
30-second averages of each therapy parameter related to
the therapy provided to the patient
Waveform data including pressure, flow, volume, and total
leak
Patient monitors:
-
CO
monitor data
2
-
Oximeter data
-
FiO
monitor data
2
Export details about the patient's therapy to a storage device connected through
Bluetooth. Data is identified by the device serial number.
Data type
Changes to the ventilator, alarms, prescriptions, accessories
and oxygen
30-second averages of each therapy parameter related to
the therapy provided to the patient
Waveform data including pressure, flow, volume, and total
leak
Patient monitors:
-
CO
monitor data
2
-
Oximeter data
-
FiO
monitor data
2
Export the alarm and event logs. If two storage devices are connected, the logs are
saved to the device that was connected first.
Installs prescriptions that were previously exported.
Ensure the version number that you are installing is higher than the current software
version. After installing the update, the system automatically restarts.
Bluetooth Data Transfer icon
Data Transfer
Instructions for Use | Device Options
Data stored
6 months
31 days
Data stored
6 months
31 days
69

Advertisement

Table of Contents
loading

Table of Contents