Setting The Alarm Volume; Responding To An Alarm; Resetting Alarms - Philips Trilogy Evo Universal Instructions For Use Manual

Table of Contents

Advertisement

Setting the Alarm Volume

Warning:
Ensure the alarm volume is set loud enough for the caregiver to hear it. Consider the use of a remote alarm. If you do use a remote alarm, test it
before starting ventilation.
To set the alarm volume:
1.
In the menu bar, tap the Options icon.
2.
In the Options window, tap Device Settings.
3.
In the Device Settings window, tap Alarm Volume.
4.
In the setting dialog box, make your selection.
5.
On the Alarm Volume dialog box, select the volume you want and then tap the Accept checkmark.

Responding to an Alarm

Warning
Visually monitor the patient and ventilator at all times during an Alarm Silence period. Allowing alarms to continue without intervention may result
in harm to the patient.
When an alarm occurs:
1.
Ensure the patient has adequate ventilation and oxygen. If required, provide an alternate method of ventilation.
2.
Tap the Alarm List to view all alarms and messages. If you see the help icon
3.
Press the Alarm Silence button on the device to pause all audible alarms for 2 minutes.
4.
Take action to resolve the alarm. For help on specific alarms, see the "Alarms and System Messages" section.

Resetting Alarms

The Alarm Reset icon resets all active and resolved alarms.
When an alarm-triggering condition is no longer present, the alarm status changes to Resolved and the alarm or message remains in the list. Tap
the Alarm Reset icon to reset the alarm list.
Instructions for Use | Alarms and System Messages
Resetting Alarms
, you can tap it for more information.