Therapy Actions
Oxygen Flush
D
ESCRIPTION
This feature requires the oxygen blender. When active, the device delivers 100% oxygen for two minutes. This feature functions independent of
any oxygen blending setting. During an oxygen flush, the High FiO
W
O
F
ORKING WITH
XYGEN
To start an oxygen flush:
Tap 100% O
in the status bar and then tap Start. A timer appears that counts down
2
the two minutes.
To stop an oxygen flush:
Tap 100% O
in the status bar and then tap Stop.
2
A
T
M
PPLICABLE
HERAPY
ODES
All
Manual Breath
D
ESCRIPTION
Delivers a breath based on the current therapy mode settings.
To deliver a manual breath:
Tap the Manual Breath button in the status bar and then tap Start.
A
T
M
PPLICABLE
HERAPY
ODES
•
A/C-PC
•
A/C-VC
•
PSV
•
S/T
•
SIMV-PC
•
SIMV-VC
Therapy Control Settings
Therapy control settings can be interdependent. For guidance, see the previous therapy mode descriptions.
Setting Name
Setting Range/Increment
Adult patient type:
•
All circuits but passive: 0-60 cmH
•
Passive circuit: 0-57 cmH
•
Increments: 1 cmH
Backup Pressure
Support
Pediatric patient type:
•
All circuits: 0-30 cmH
•
Increments: 1 cmH
Infant patient type:
LUSH
O
2
O
2
O
2
O
2
O
2
Therapy Control Settings
Instructions for Use | Therapy Modes and Controls
alarm is disabled.