Monitored parameters and alarms that use the flow measurement such as tidal volume will not
alarm
function.
Monitored parameters and alarms that use the pressure measurement will continue to function.
E
F
S
R
XTERNAL
LOW
ENSOR
Priority
High
Why it occurs
The external flow sensor is connected backwards during active therapy.
What to do
Check the sensor position. The arrow direction should align with the air delivered to the patient.
Requirements
Active flow or dual limb circuit
Device
In volume modes, device will not accurately control volume. The peak inspiratory pressure will
performance
continue to increase.
during active
In pressure modes, therapy will be unaffected.
alarm
All monitored parameters and alarms will continue to function.
A
E
V
CTIVE
XHALATION
ALVE
Priority
High
Why it occurs
The active exhalation valve is stuck closed.
What to do
Requirements
Active flow or dual limb circuit
Device
The alarm is automatically resolved and the device continues to operate.
performance
during active
alarm
C
A
E
HECK
CTIVE
XHALATION
Priority
High
Why it occurs
The active-exhalation pressure control line is not connected, becomes disconnected, or contains water
droplets that affect the active exhalation valve line pressure reading.
What to do
Inspect the line. Empty or replace it if necessary.
Requirements
Active flow or dual limb circuit
Device
Therapy delivery will be compromised.
performance
All monitored parameters and alarms will continue to function.
during active
alarm
P
P
L
ROXIMAL
RESSURE
INE
Priority
High
Why it occurs
The proximal pressure line is not connected.
What to do
Verify the proximal pressure line is connected properly at both ends and the line is clean and
untangled.
Ensure the main circuit is connected properly and does not have large leaks.
Ensure the exhalation valve is intact.
Requirements
Dual limb, active PAP, or active flow circuit
Device
The alarm is automatically resolved when the proximal pressure line is connected properly.
performance
during active
The device continues to operate.
alarm
O
R
XYGEN
EGULATION
Priority
High
Why it occurs
The flow measured FiO
What to do
EVERSED
F
AILED
For a single-limb active circuit, check all connections to the valve and check that the valve is clear.
For a dual limb circuit, check that the valve is clear..
P
C
L
RESSURE
ONTROL
INE
D
ISCONNECTED
is not within 10% ± 3% of the FiO
2
Verify the oxygen blender is connected properly.
Alarms and System Messages
Instructions for Use | Alarms and System Messages
setting.
2
47