Alarm And Event Log - Philips Trilogy Evo Universal Instructions For Use Manual

Table of Contents

Advertisement

E
CO
T
2
1.
Set the High EtCO
2
2.
Confirm that the High EtCO
3.
Restore the alarm limit.
4.
Set the Low EtCO
2
5.
Confirm the Low EtCO
6.
Restore the alarm limit.
P
R
ULSE
ATE
1.
Set the High Pulse Rate alarm limit below the measured Pulse Rate value.
2.
Confirm that the High Pulse Rate alarm activates.
3.
Restore the alarm limit.
4.
Set the Low Pulse Rate alarm limit above the measured Pulse Rate value.
5.
Confirm the Low Pulse Rate alarm activates.
6.
Restore the alarm limit.
S
O
P
2
1.
Set the High SpO
alarm limit below the measured SpO
2
2.
Confirm that the High SpO
3.
Restore the alarm limit.
4.
Set the Low SpO
alarm limit above the measured SpO
2
5.
Confirm the Low SpO
6.
Restore the alarm limit.

Alarm and Event Log

The Alarm and Event Log is a record of all events related to the device and to therapy. The log shows each event, when it occurred, and a brief
description. Information is retained even when you shut off the device or when power is lost. The log stores the most recent 6-months of
information with the exception of the event log, which stores the most recent 10,000 records. Older records are overwritten.
To access the Alarm and Event Log:
1.
In the menu bar, tap the Options icon.
2.
In the Options window, tap Alarm & Event Log.
alarm limit below the measured EtCO
alarm activates.
2
alarm limit above the measured EtCO
alarm activates.
2
alarm activates.
2
alarm activates.
2
Alarm and Event Log
Instructions for Use | Alarms and System Messages
value.
2
value.
2
value.
2
value.
2
63