A/C-VC: Assisted/Control-Volume Control
D
:
ESCRIPTION
The A/C-VC mode provides volume-controlled mandatory and assist breaths. When the Trigger Type is set to Off, both triggering and cycling are
performed by the ventilator. When the Trigger Type is not Off, then the trigger can be performed by the vent or patient and the cycle is always
performed by the ventilator. To deliver the set volume in the set time, the ventilator alters the flow rate. The flow pattern setting defines the shape
of the flow delivery pattern.
S
:
ETTINGS
Setting Name
Description
Tidal Volume
Set inspiratory volume
PEEP
Positive end expiratory pressure
Inspiratory Time
Length of the inspiratory phase
Breath Rate
Minimum rate of mandatory breaths per minute
Flow Pattern
Sets the shape of the waveform as a ramp or square
Trigger Type
Auto-Trak
Sensitive Auto-Trak (passive circuits only)
Flow Trigger (Passive, Active PAP, Active Flow, or Dual Limb circuits)
Off
Flow Trigger
This control is available when the trigger type is Flow Trigger. The flow trigger initiates when the
Sensitivity
patient's inspiratory effort creates a flow equal to or greater than the flow trigger sensitivity setting.
Flow Cycle
This control is available when the trigger type is Flow Trigger.
Sensitivity
As flow begins to decrease during inspiration, if the patient flow is less than the flow cycle set point, the
device cycles to expiration.
S
:
ETTABLE ALARMS
Circuit disconnected
High tidal volume
Low tidal volume
High minute ventilation
Low minute ventilation
High respiratory rate
Low respiratory rate
High inspiratory pressure
Low inspiratory pressure
I
LLUSTRATION
Instructions for Use | Therapy Modes and Controls
Control Modes
20