Specifications; Temperature; Pressure - Fieldpiece SMAN SM480V Operator's Manual

Refrigerant manifold + micron gauge (4 port)
Table of Contents

Advertisement

Specifications

Display: LCD (5 inches diagonal)
Backlight: Blue (adjustable duration)
Low Battery Indication:
drops below the operating level.
Over Range Display: OL for pressure, - - - - for temperature
Auto Power Off: 30 minutes of inactivity (adjustable)
Maximum Manifold Pressure: 800 Psig (5500 kPa)
Battery Type: 6 x AA alkaline
Battery Life: 350 hours typical
(without vacuum, backlight, and wireless)
Radio Frequency: 2.4 GHz
Wireless Range: 350 feet (106 meters) line of sight.
Distance decreases through obstructions.
Data Port: Micro USB (for extracting data logs or updating firmware)
Operating Environment: 32°F to 122°F (0°C to 50°C) at <75% RH
Storage Environment: -4°F to 140°F (-20°C to 60°C) at <80% RH
(with battery removed)
Temperature Coefficient: 0.1 x (specified accuracy) per °C (0°C to
18°C, 28°C to 50°C), per 1.8°F (32°F to 64°F, 82°F to 122°F)
Weight: 4.03 lbs (1.83 kg)
Water Resistant: Designed to IP54
US Patent: www.fieldpiece.com/patents
Refrigerants: New refrigerants are continually being added so be
sure to visit www.fieldpiece.com for the latest firmware.
R11
R115
R290
R407A
R12
R116
R401A
R407C
R13
R123
R401B
R407F
R22
R124
R402A
R408A
R23
R125
R402B
R409A
R32
R134A
R403B
R410A
R113
R236FA
R404A
R413A
R114
R245FA
R406A
R414A
is displayed when the battery voltage
R414B
R422B
R448A
R500
R416A
R422C
R449A
R501
R417A
R422D
R450A
R502
R417C
R424A
R452A
R503
R420A
R427A
R452B
R507A
R421A
R428A
R453A
R508B
R421B
R434A
R454B
R513A
R422A
R438A
R458A
R600
*Maximum pressure: 580 Psig (4000 kPa)

Temperature

Sensor Type: Type K thermocouple
(nickel chromium/nickel aluminum)
Jack Type: (3) Type K thermocouple
Range: -50°F to 257°F (-46°C to 125°C), limited by the thermocouple
specification. Display range is -95°F to 999.9°F (-70°C to 537.0°C).
Resolution: 0.1°F (0.1°C)
Accuracy: Shown accuracies are after field calibration.
±(1.0°F) -95°F to 200°F, ±(2.0°F) 200°F to 999.9°F;
±(0.5°C) -70°C to 93°C, ±(1.0°C) 93°C to 537.0°C

Pressure

Sensor Type: Absolute pressure sensors
Port Type: (1) 3/8" and (3) 1/4" standard NPT male flare fittings
Pressure Range and Units: 580 Psig (English), 40.00 Bar (Metric),
4.000 MPa (Metric), and 4000 kPa (Metric)
Negative Pressure Range and Units:
29 inHg (English), 74 cmHg (Metric)
Resolution: 0.1 Psig; 0.01 Bar; 0.001 MPa; 1 kPa; 0.1 inHg; 1 cmHg
Negative Pressure Accuracy:
29 inHg to 0 inHg: ±0.2 inHg;
74 cmHg to 0 cmHg: ±1 cmHg
Pressure Accuracy:
0 Psig to 200 Psig: ±1 Psig;
200 Psig to 580 Psig: ±(0.3% of reading + 1 Psig);
0 Bar to 13.78 Bar ±0.07 Bar;
R600A
13.78 Bar to 40.00 Bar: ±(0.3% of reading + 0.07 Bar);
R601
0 MPa to 1.378 MPa: ±0.007 MPa;
R601A
1.378 MPa to 4.000 MPa: ±(0.3% of reading + 0.007 MPa);
R744*
R1233ZD
0 kPa to 1378 kPa: ±7 kPa;
R1234YF
1378 kPa to 4000 kPa: ±(0.3% of reading + 7 kPa)
R1234ZE
37

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sman sm380v