Page 3
Table of Content (TOC) General Features Instrument Setup Setting time/date of instrument KI2400/KI2800 series light source overview & keypad layout Changing emitters’ output power level Getting started Activating / deactivating wavelengths for Limited Getting to know the light source model styles...
1. General Features Long battery life (up to 90 hours) Up to 6 mixed LED, Laser & VFL sources • • Single port for MM, Single port for SM • Autotetes capability • VisiTester option for easy active fiber channel •...
Page 5
2. KI2400/KI2800 series light source overview & keypad layout Dust Cover Remove Connector Lanyard holder Display backlight On/Off USB port for power/battery charging Battery charge indicator Menu [F1 ]~[F4] [-], [+] Toggle Centre Go back to TOC...
3. Getting started 3.1. Getting to know the light source model styles 3.2. Selecting the appropriate Launch Test Cord 3.3. Install/uninstall optical connector adaptor 3.4. Install/uninstall batteries 3.5. Enable/disable battery charging 3.6. Switching on/off instrument 3.7. Switching on/off LCD display backlight Go back to TOC...
Page 7
3.1. Getting to know the light source model styles [Go back to Getting Started section] There are two model styles: I. KI2400 Series - preferred by telcos & those requiring higher stability Ultra Stable laser light source • Single port •...
Page 8
To avoid damaging the instrument’s connector tip, it is important to use a Test Cord (not supplied) with the correct connector type that matches with that on the instrument. The instrument is supplied with SC/SC hybrid adaptor. Choose to order other required adaptor types from Kingfisher’s website. Go back to TOC...
Page 9
3.3. Install/uninstall optical connector adaptor [Go back to Getting Started section] Press and hold down Release Button with ➀ one hand to unlatch. Adaptor slot ❷ ❸ face out Pull out existing adaptor with the other ➁ hand. Push in a new adaptor until it latches on ➂...
Page 10
3.4. Install/uninstall batteries [Go back to Getting Started section] To install or uninstall batteries in the battery compartment, Unclip Battery Compartment Cover at rear of instrument. ❶ Insert/remove batteries (take note of the battery +Ve terminal spring orientation of different instrument models, see images below). ❷...
Page 11
3.5. Enable/disable battery charging [Go back to Getting Started section] To enable battery charging, the supplied 0.1” pitch shorting-pin must be properly installed in battery compartment Instrument supplied in this configuration i.e. battery charging is Disabled: Supplied 0.1” pitch shorting-pin Insert the supplied shorting-pin as shown to Enable battery charging: Supplied 0.1”...
Page 12
3.6. Switch on/off instrument [Go back to Getting Started section] To switch on instrument, press the green, [On/Off ] button. To switch off instrument: press the green, [On/Off ] button again. Note: the instrument will switch off automatically 10 minutes after it was switched on.
Page 13
3.7. Switch on/off LCD display backlight [Go back to Getting Started section] To switch on backlight of the display, press button with the illumination symbol, To switch backlight off, press that button again. Go back to TOC...
Page 14
4. Instrument Menu Structure Top level Menu Auto Source Source 2’nd level Menu Mode Home *Set Ref Home Turn off Stop Select CW Change Function emitter, Autotest & Start or tone output return to top return to top Autotest power level modes level menu level menu...
5. Instrument Operations The light source can be operated in the modes below. 5.1. Normal Autotest Operation Mode: • Automatically toggle all available wavelengths. • Preferred mode for loss testing as testing time is greatly reduced. • Minimises error as meter always displays the correct wavelength. 5.2.
Page 16
5.1. Normal Autotest Operation Mode: [Go back to Instrument Operations section] A. To start Autotest for light source with 1 to 4 wavelengths, At top level menu, press [F4/Auto] (Fig. 1). • All available wavelengths will be displayed one at a time in sequence. See Fig.3 on the right for an example with 2 active wavelengths.
Page 17
5.2. Limited Wavelength Autotest Operation Mode: [Go back to Instrument Operations section] To start Limited wavelength Autotest, At top level menu, press & hold [Toggle Centre] then press [F4/Auto]. Per m Per m Aut o Aut o Hom e Hom e Programmed out power level •...
5.3. Manual Operation Mode: [Go back to Instrument Operations section] ➀ At top level menu, press ➁ Press [- or+] repeatedly to select other wavelengths or turn off emitter in the sequence shown below. [F1/Source] to turn on emitter. The default wavelength will be •...
Page 19
5.4. Tone Operation Mode: [Go back to Instrument Operations section] To work with Tone mode at a wavelength, ➂ Press [F3/Mode] ➀ At top level ➁ Select the To exit Tone mode & to go into Tone menu, press desired return to top level menu, mode.
Page 20
5.4. Tone Operation Mode (continue): [Go back to Instrument Operations section] In Tone mode, to select from 1 of the 15 available test/FiberID tones, 270, 1000, 2000, id01, id02…id12, Press [F1/Source] repeatedly to scroll to the desired test tone in the sequence shown below. Sou rc e Sou rc e Sou rc e...
6. Instrument Setup: 6.1. Setting time/date of instrument 6.2. Changing emitter’s output power level 6.3. Activating/deactivating wavelengths for Limited Wavelength Autotest operation Go back to TOC...
Page 22
6.1. Setting date/time of instrument: [Go back to Instrument Setup section] To enter time/date setting mode: ➁ To select date/time item for setting: ➀ ➂ To modify selected time/date item: Press [Toggle Centre] [F3]. When instrument is OFF, press and Press [+].
Page 23
6.2. Changing emitters’ output power level: [Go back to Instrument Setup section] [Go back to Normal Autotest Operation Mode section] [Go back to Limited Wavelength Autotest Operation Mode section] [Go back to Manual Operation Mode section] To adjust output power level of a turned on emitter, Press and hold Ref], then press or+].
Page 24
6.3. Activating / deactivating wavelengths for Limited Wavelength Autotest operation [Go back to Instrument Setup section] [Go back to Limited Wavelength Autotest Operation Mode section] To deactivate/activate wavelength when emitter is turned on, Whilst [Toggle Centre] is being hold, the status “out”...
7. Instrument Firmware The KI2000 series Firmware can be end-user upgraded. To update Firmware: Instrument Firmware must be r0.12 or higher. • KI2000 USB device driver software must be installed. • 7.1. Checking Firmware Version 7.2. Firmware Upgrading Procedure Go back to TOC...
7.1. Checking firmware version [Go back to Instrument Firmware section] The instrument firmware version can be checked during instrument switch-on. When instrument if off, hold down [F4] [On/Off ] at the same time: The instrument will switch on, and serial number and firmware version is displayed whilst [F4] is held.
7.2. Firmware upgrade procedure [Go back to Instrument Firmware section] Download and extract the Firmware Update program from Kingfisher web site. ❶ Connect instrument to computer. ❷ Run Firmware Update program (KI2000 programmer vx.xx.exe) downloadable free from Kingfisher website. ❸...
8. Instrument care Keep the instrument in a carry case during storage and transport • Use only high quality batteries. • For prolonged storage remove batteries. • The instrument is resistant to normal dust and moisture, however it is not waterproof. •...
Need help?