3. Preparation
3.4. Calibrating the transmitter

Wiring for calibration

ATTENTION
When two wires will be installed in one screw terminal, both wires must be
the same size (gauge) and construction (solid or stranded).
SINGLE THERMOCOUPLE
CONNECTION
T/C or mV
1
2
(PV1)
3
TB1
4 JUMPER WIRES
T/C MAY BE GROUNDED TYPE
THREE THERMOCOUPLE
TB1
T/C or mV
1
2
(PV1)
3
T/C or mV
(PV3)
2 JUMPER WIRES
ONLY ONE T/C MAY BE GROUNDED TYPE
TWO THREE-WIRE RTD's
TB1
RTD
1
or
Ohms
2
(PV1)
3
T/C & SINGLE THREE -WIRE RTD
TB1
T/C or mV
1
2
(PV1)
3
T/C MUST BE FLOATING (UNGROUNDED) TYPE
14
OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual
4
5
6
TB2
CONNECTION
TB2
4
T/C or mV
5
6
TB2
4
5
6
TB2
4
5
6
TWO THERMOCOUPLE
1
T/C or mV
2
3
(PV1)
3 JUMPER WIRES
ONLY ONE T/C MAY BE GROUNDED TYPE
SINGLE THREE -WIRE RTD
RTD
or
1
(PV2)
Ohms
2
(PV1)
3
SINGLE THREE -WIRE RTD & T/C
RTD
1
or
RTD
Ohms
or
2
(PV1)
Ohms
3
(PV2)
T/C MUST BE FLOATING (UNGROUNDED) TYPE
SINGLE TWO-WIRE RTD
RTD
RTD
or
1
or
Ohms
Ohms
2
(PV1)
(PV2)
3
CONNECTION
4
5
6
TB1
TB2
TB1
TB2
4
5
6
TB1
TB2
4
5
6
TB1
TB2
4
5
6
JUMPER WIRE
T/C or mV
(PV2)
T/C or mV
(PV2)
Rev. 3
6/24/08