Welch Allyn Acuity Directions For Use Manual page 134

Central monitoring systems
Hide thumbs Also See for Acuity:
Table of Contents

Advertisement

130
Chapter 7 Move a monitored patient
To move a patient from a Propaq LT monitor to a different monitor
1.
On the Propaq LT monitor, press the arrow keys to underline
the top of the screen, and then press
2. Select Acuity menu.
3. Select END TELE.
The monitor disconnects from the Acuity System.
4. Turn off the monitor.
5. Select Delete & Shut Down.
The trend information for this patient is erased from the monitor.
6. Connect the patient to a new monitor, and connect the monitor to the Acuity System.
At the central station, the patient's yellow icon appears in an unconfirmed state on
the map.
7 .
Enter or confirm the patient information.
For instructions, see
on page 41.
To move a patient from a wireless Propaq CS monitor to a different monitor
1.
On the Propaq CS monitor, touch NET OFF .
2. Turn off the monitor.
The trend information for this patient is erased from the monitor.
3. Connect the patient to a new monitor and connect the monitor to the Acuity System.
At the central station, the patient's yellow icon appears in an unconfirmed state on
the map.
4. Enter or confirm the patient information.
For instructions, see
on page 41.
to select it.
"Enter and confirm a patient's ID or room at the central station"
"Enter and confirm a patient's ID or room at the central station"
Welch Allyn Acuity Central Monitoring System
at

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mobile acuity lt

Table of Contents