Chapter 3. Setting Up the DM5E

3.6 Locking and Unlocking Instrument the Controls

The DM5E allows operators to lockout specific functions by using the lockout display, which is only accessible from the measurement
screen. Follow the instructions in Figure 13 on page 33 to access this display and lockout (or enable) some or all of the function keys.
Note: The DATA RECORDER and DUAL MULTI functions are optional and are only available with advanced instrument versions.
However, an upgrade of the basic version is possible by purchasing a license key.
In the Lockout menu, the operator can set the probe zeroing and calibration mode (CAL/ZERO MODE) by selecting 1pt, 2pt or AUTO. One
of the following calibration reminder styles can also be selected:
REMIND TIME - The calibration reminder is displayed after a preset elapsed time from the last thickness calibration. The setting
indicates the preset time value.
CAL READINGS - The calibration reminder is displayed after the user presses the SEND function the number of times indicated by the
setting value.
CAL POWER ON - The calibration reminder is displayed each time that the instrument is powered ON when the value is set to ON.