Using the Z Vent Ventilator

Initial Operation with Default Parameter Settings

Following power on, the ventilator goes through a Self-Check (a set of system tests and
checks). If the Self-Check passes, the LED array turns green and the Start Menu displays,
indicating that the ventilator is operational.
The Start Menu enables the user to choose between predefined ventilator parameter settings
(for adult patients, pediatric patients, patients requiring Mask CPAP), a previously saved set of
Custom parameter settings, or the parameter settings last used during ventilator operation. The
Start Menu choices include:
Choice
Adult
Pediatric
Mask CPAP
Custom
Last Settings
Warning!
Do not connect patient to the ventilator while the Start Menu is active.

Default Parameter Settings for Adult, Pediatric and Mask CPAP

The default settings for Adult, Pediatric, and Mask CPAP are as follows:

Adult Default Parameter Setting Values

Parameter
Mode
BPM
I:E
VT
PEEP
PIP Limit
FIO
2
4-2
Description
Preset ventilation parameter settings for adult patients.
Preset ventilation parameter settings for pediatric patients.
Preset ventilation parameter settings for Mask CPAP (Continuous
Positive Airway Pressure) ventilation.
Ventilation parameter settings previously saved by a user.
The settings enabled on the ventilator during its last use (but not
saved as Custom Settings by the user).
Default Setting Value
AC (V)
12
1:3
450
5
35
21
www.zoll.com
9650-002360-01 Rev. A