Vt Parameter Window Reference

The following table provides the options and ranges for the Vt window parameters:
Vt Parameter Window
Primary Parameter
Secondary Parameters
Alarm Limits
Vt Context Menu
Tubing Compliance (CT) Default: Off
Adult
Infant
Compliance Volume (ml) (Measured value)

PIP Parameter Window Options

The PIP (Peak Inspiratory Pressure) parameter window displays and/or controls airway
pressure. The breath target determines if the PIP value is a measurement or a control setting
(solid text). When PIP is a control setting, the user can adjust PIP from 10 to 80 cm H
Note:
During pressure targeted ventilation, the PIP parameter is displayed as solid text.
During volume targeted ventilation, the PIP primary parameter is a measurement and displays
as outlined text.
Positive End Expiratory Pressure (PEEP) is the baseline pressure maintained above
atmospheric pressure at the end of exhalation to prevent alveolar collapse and improve gas
exchange.
For BL mode, the ventilator provides noninvasive ventilation with the ability to manage the
patient by adjusting the IPAP and EPAP parameters. IPAP is adjustable from 6 to 60 cm H
EPAP is adjustable from 3 to 30 cm H

Secondary Parameters and Alarm Thresholds/Limits

In pressure targeted ventilation mode, the following secondary parameters are available.
High PIP Alarm Limit
Low PIP Alarm Limit
PEEP
PS (Available for SIMV(P), CPAP, and BL)
9650-002360-01 Rev. A
Vt
ml
None
High Vt
Low Vt
Default: 1.60
Default: 0.50
When the user attempts to set a PIP value greater than 60 cm H
displays prompting for a confirmation before values above 60 cm H
ZOLL Z Vent Ventilator Operator's Guide
Options/Range Availability/Notes
50 to 2000
50 to 2000, Off
5 to 500, Off
OFF, Adult,
Infant
0 to 3.50
0 to 2.00
0 to 349
O.
2
Using the Z Vent Ventilator
Volume Target: Control Setting
Pressure Target: Measured
Limits are not present during
volume targeted ventilation
Auxiliary Box
User entered values are not
retained when the ventilator is
turned OFF
0.
2
O, a popup message
2
0 are available.
2
O,
2
4-17