Indicators; Display Screen - ZOLL Z Vent Operator's Manual

Table of Contents

Advertisement

Product Overview
Control
Selection Dial
Mute/Cancel button
Accept/Select button
Manual Breath Button/
Plateau Pressure

Indicators

The ventilator's indicators consist of the following:
Indicator
LCD Display
LED Array

Display Screen

The ventilator's display screen has four functional areas as shown in Figure 2-3:
Alarm Message Center/Waveform Window
Parameter Windows
Shared Icon Area
Auxiliary Parameter Boxes.
These functional areas are discussed in the following sections.
2-4
Function
Enables the user to set values for a chosen (highlighted) Primary
Parameter, Secondary Parameter, Context Menu item, and Menu item.
Values accelerate with speed of turning.
The Mute/Cancel button mutes the audible alarm allowing the user time
to change parameters. It can also be used to cancel parameter entries.
The Accept/Select button allows the user to accept parameter value
settings, acknowledge popup messages, and accept a menu choices
Enables the user to deliver a manual breath and measure Plateau
Pressure
Note: Plateau Pressure is an optional ventilator control.
Description
Displays settings, patient data, and alarm information.
Indicates operational status (Red, Yellow, or Green).
Figure 2-3 Display Screen Functional Areas
www.zoll.com
9650-002360-01 Rev. A

Advertisement

Table of Contents
loading