Place the leading edge of the paper onto the metal bar and push it into the unit under the
roller bar. The instrument will automatically feed the paper through once you have inserted
it far enough.
Close the Paper Access Door, being sure the paper is sticking out through the slot in the
door. Tear off any excess paper if desired.
Note: The printer paper can be advanced by selecting ADVANCE PAPER (2nd Operator
Button from the top) in the results menu or by selecting ADVANCE PAPER (3rd
Operator Button from the top) in the FDT MAIN MENU.
FREQUENCY DOUBLING TECHNOLOGY MAIN
RUN PATIENT TESTS
RUN DEMONSTRATION
ADVANCE PAPER
UTILITIES MENU
Note: Use only an appropriate heat-sensitive printer paper designed to be used with the
Seiko Instruments 5000 series printer inside the instrument (reverse wound rolls only) or
the print quality may be degraded, the printer life may be shortened, and the instrument
warranty voided. Appropriate paper may be ordered through any authorized
representative. Refer to the Replacement Parts and Accessories section of this manual
for part number and ordering information.
Note: Because the printer paper is thermally activated, no printing will appear on the
paper if it is inserted backwards.
Note: The printer paper is thermally activated, so it must be stored in a cool, dry, dark
location to prevent exposure and degraded performance over time.
Note: Do not use transparent adhesive tape on printed portions of the printout, as those
portions of the printout will then fade.
Note: Do not store the printed side of the printout in contact with plastic folders or
sheets, as those portions of the printout will then fade.
Note: Do not allow any cleaning or disinfection solutions or other liquids to come into
contact with the printer paper. Degraded print quality or printer damage may occur for
new printouts and degraded printouts of previously printed results may occur (especially
with Isopropyl alcohol).