DIAGNOSTIC TOOL
SMART PHONE APPLICATION
You can connect to this chair using an Android or iOS based smart phone and the DCI Edge Diagnostic Tool
App. This application provides the following features:
1. View status information of the chair in its current state along with troubleshooting tips for error conditions.
2. Download a history log of past actions and conditions.
3. Share the history with the DCI Edge Technical Service Team for additional assistance.
4. Gain quick access to other DCI Edge support materials available online.
5. Quickly connect with the DCI Edge Technical Service Team by phone or by electronic message.
CONNECT VIA BLUETOOTH
Getting started: install the DCI Edge Diagnostic
1
Tool onto your smart phone from the GooglePlay
Store or Apple AppStore. Once you have installed the app
on your smart phone, open the app and choose Devices
from the main menu.
Make the chair discoverable: Using a keypad or foot-
2
switch connected to the chair, press and hold the '0' and
'1' keys simultaneously for 5 seconds until you hear a chime
from the chair.
Connect: Choose Devices from the main menu (Fig. 1).
3
From the Devices Screen, tap on the device listed to
connect to it (Fig 2). To avoid confusion, do not make more
than one chair discoverable at a time.
View Status: Once connected, you will be placed
4
into the Status Section of the app (Fig. 3) where you
can view the codes that are currently being displayed on
the Control Board of the chair along with the definition of
the code. A graphical display will also indicate any current
actions you are making in real time. Scroll through the
troubleshooting section at the bottom of the screen if the
current conditions require a corrective action.
View Log: Tap Log in the upper left of the screen to
5
enter the Log Section of the app (Fig. 4) where you
can view the Log Entries in real time as you operate the
chair. You can also download the log history of the chair
for the last 500 log entries, and you can tap the Share icon
to email that log information to DCI Edge Technical Service
Team for additional assistance. The Log Help window
includes additional information for understanding the log
entries and viewing the list of troubleshooting tips.