Figure 6-38 Wiring Diagram: Electrical Enclosure - GE Giraffe Service Manual

Hide thumbs Also See for Giraffe:
Table of Contents

Advertisement

Chapter 6: Illustrated Parts
6600-0746-700
Ebase / Hood
Transformer
Standby Switch
1
6600-0227-850
2
6600-0800-700
Outlets
6600-0583-600 (X3)
158
Ebase Transformer
Configuration Plug Group 1
Standby Switch
BATTERY
3
BT1
6600-1024-600
4
Power Supply
Power Supply
6600-0708-700
6600-0221-850
+5 V
+5V
PHASE
+5V
GND
NEUTRAL
GND
GND
AC GND
+12V
+12 V
Mains Outlets
P
G
N
1
P
G
1
N
6600-0562-603 X2
P
Circuit Breakers 3.5A.
G
N
To J51 on
Relay Board

Figure 6-38 Wiring Diagram: Electrical Enclosure

6600-0356-000 103
Sec A 14.6vac
Sec A 14.6vac
Sec B 14.6vac
Sec B 14.6vac
6600-0748-702
6600-0748-701
230V
Configuration Plug Group 1
V charge out
V charge in
V battery
GND
6600-0707-700
red
white
blue
green
black
brown
orange
yellow
red
black
2
3
1
4
2
2
Power Switch/
Circuit Breaker
6600-1014-601 (9 Amp, 230V)
6600-1014-602 (12 Amp, 115V)
brown
6600-0731-700
blue
includes wiring to outlets
red
To J44 on
white
violet
Relay Board
black
To J52 on
Phase
black
Neutral
white
Relay Board
115V
To J43 on
red
black
Relay Board
red
black
To J47 on
Phase in
red
Phase out
black
Relay Board
To J41 on
Relay Board
To J48 on
Relay Board
P
Line Filter
G
N
6600-1006-600
120VAC into relay board and out
to power supply. If in good and
out missing check fuses on relay
board.
© 2001 by Datex-Ohmeda, Inc.. All rights reserved.
When checking for the
secondary voltage 5
and 12 VDC from the
power supply to the
relay board J 41, make
sure the power cord is
connected and both
the breaker on the
back and the switch
on the front panel are
in the ON position.