Figure 4-5 First Service Screen: Diagnostics - GE Giraffe Service Manual

Hide thumbs Also See for Giraffe:
Table of Contents

Advertisement

Item
Description
Brings up scale calibration screen. Refer to "Chapter 3: Calibration and Maintenance" for
Cal Scale
information on how this screen is used.
A number of diagnostic readings appear on the right side of the service screens.
Item
Description
HFS
High fan speed. Should be 1500 + 100 (measured at power up only)
LFS
Low fan speed. Should be 1000 + 100 (measured at power up only)
RH
Relative Humidity. % humidity read in the patient chamber
SR
Scale counts raw
SC
Scale counts corrected (1 count = 1 gram)
Air display temperature. Temperature read by the first thermistor in the compartment air probe.
ADT
Should be + 0.3ºC of ACT temperature.
Temperature read by second thermistor in the compartment air probe. Should be + 0.3ºC of ADT
ACT
temperature.
P11
Reading from the first thermistor in patient jack 1. Should be + 0.5ºC of P12 temperature.
P12
Reading from the second thermistor in patient jack 1. Should be + 0.5ºC of P11 temperature.
P21
Reading from the first thermistor in patient jack 2. Should be + 0.5ºC of P22 temperature.
P22
Reading from the second thermistor in patient jack 2. Should be + 0.5ºC of P21 temperature.
Heat sink sensor resistance. Should be approximately 20000 ohms @ 25ºC. Refer to the Tips section
HSP
4.6.7 for resistance verses temperature values.
LV1
Line voltage in first mains circuit.
LF
60Hz or 50Hz
MC
Motor current. Shows current drawn by the e-base motor
TV
Thermistor voltage. Voltage of thermistor circuits located on the mother board.
© 2001 by Datex-Ohmeda, Inc.. All rights reserved.

Figure 4-5 First Service Screen: Diagnostics

Chapter 4: Troubleshooting
6600-0356-000 103