4.3.3 "Baby Hot" or "Baby Cold" Alarm

Cause or Circumstances
Baby Hot or Baby Cold
alarm activated while unit
is in Air Mode.
MANUAL TEMP alarm is
set as active on the nurse
setup screen or the service
screen.

4.3.4 "Baby Mode Disabled" Alarm

Cause or Circumstances
Baby mode has been
disabled in the service
screen and baby mode
button is pressed.

4.3.5 "Baby Probe 1 (or 2) Failure" Alarm

Cause or Circumstances
Probe thermistors show
difference > 0.5C.

4.3.6 "Bad Membrane Switch" Alarm

Cause or Circumstances
All
© 2001 by Datex-Ohmeda, Inc.. All rights reserved.
Failure
None
Failure
None
Failure
Defective patient probe
Defective patient probe
jack wiring
Failure
Defective membrane
switch panel
Action
On service screen or nurse setup screen, select OFF
for the MAN TEMP ALARM setting.
Action
Press alarm silence button to deactivate the alarm.
If you wish to use the baby mode, re-enable it in
service screen.
Action
Replace patient skin probe.
Plug patient probe into probe jack 1. In service
mode, note the difference between readings from
P11 and P12. Plug the same probe into probe jack 2
and note the difference between readings from P21
and P22. If they are greater there is resistance in the
probe jack harness. Replace the harness.
Action
Test switches in service mode. Replace membrane
switch panel.
6600-0356-000 103