Appendix B: Specifications
B.4 Operating Environment
Temperature
Humidity
Pressure
B.5 Storage Conditions
Temperature
Humidity
Pressure
B.6 User Control Settings
Patient control temperature
Air control temperature
Humidity: Servo- % relative humidity
B.7 Alarms
High Air Temp
Low Air Temp
Baby Hot
Baby Cold
Fan Failure
Air Temp >38ºC
Air Temp >40ºC
Air Probe Failure
Air Probe Disconnect
Power failure
System failure
Add Water
Scale: Weight Exceeds Maximum
Scale: Scale Failure
* Can be re-set on the Service screen to 0.5ºC
170
20 to 30ºC
10 to 95% Non-condensing relative humidity
50 to 106 kPa
-25 to 60ºC
5 to 95% Non-condensing relative humidity
50 to 106 kPa
35-37.5ºC in 0.1ºC increments
20-39ºC in 0.1 increments
30-95% in 5% increments
1.5ºC over AST (air set temperature)
3.0ºC under AST
1.0ºC* over BST (baby set temperature)
1.0ºC* under BST
Failure of blower system
>38ºC for AST <37ºC
>40ºC air temperature for AST >37ºC
Compartment air probe failure
Disconnection of compartment air probe
Power switch on but no power
Non-recoverable system failure
Humidifier water level low
Greater than 8kg load (visual only)
Detectable system failure (visual only)
6600-0356-000 103