Incubator Fan/Motor/Optical Sensor; Figure 5-5 Fan Motor - GE Giraffe Service Manual

Hide thumbs Also See for Giraffe:
Table of Contents

Advertisement

Chapter 5: Repair Procedures
6. Rotate the cover slightly and push it out the right side of the unit.
7. Remove the ground wire.

5.5.2 Incubator Fan/Motor/Optical Sensor

Refer to "Figure 5-5 Fan Motor" and "Figure 5-6 Heat Sink and Fan".
Motor
Motor shaft
Motor isolators
Thermostat
Motor bracket

Figure 5-5 Fan Motor

1. Remove the rotating bed, translation deck, tilt platform and pan.
2. Remove the fan, fan hub, and fan seal from the top of the fan motor shaft.
3. Remove the chassis cover. (Refer to section 5.5.1.)
4. Disconnect the motor leads.
5. Remove the 4 screws in the motor bracket and remove the motor assembly.
6. The motor isolators pop out of the holes in the bracket.
7. To replace the optical sensor, disconnect its connector, remove the screw from the boss in the chassis that
holds it in place, and remove the assembly.
72
© 2001 by Datex-Ohmeda, Inc.. All rights reserved.
6600-0356-000 103