Chapter 4: Troubleshooting
4.4.28 System Failure 35
System failure 35 indicates: Wrong dipswitch set up.
Cause(s)
Switches set wrong
Defective control board.
4.4.29 System Failure 50
System failure 50 indicates: Incorrect display driver board System Software checksum.
Cause(s)
Defective System Software PROM on display driver
board.
Defective display driver board.
4.4.30 System Failure 51
System failure 51 indicates: Display Defective SRAM.
Cause(s)
Defective SRAM circuit on the display driver board
4.4.31 System Failure 52
System failure 52 indicates: Display No Timer2.
Cause(s)
Defective display driver board.
4.4.32 System Failure 53
System failure 53 indicates: Display Timer2 Too Fast.
Cause(s)
Defective display driver board.
48
Action(s)
Check the 8 position dipswitch on the control board. Verify that Switch 1 is on
and Switch 2 is off.
Replace control board.
Action(s)
Replace the display driver board.
Action(s)
Replace the display driver board.
Action(s)
Replace the display driver board.
6600-0356-000 103
Action(s)
Install new system software 6600-0234-850.
If the system failure persists, replace display driver
board.