Power Consumption Estimates - Kipp & Zonen LOGBOX SE Instruction Manual

Table of Contents

Advertisement

6.2 Power consumption estimates

For safe scenario's it is advised to calculate with 50-60% of the mAh capacity as indicated on the battery used. This takes
into account the self-discharge of most batteries and the possibly needed peak current for the GSM modem.
Smart digital instrument need to remain powered all the time as the power-on sequence takes time and most instruments
need to reach a temperature equilibrium. The switched on instruments will have a significant impact on the power
consumption.
A Kipp & Zonen Smart instrument on average uses 8mA on low voltages (5-9V) and 4mA on higher voltages.
Tabel updated 20-03-2018 Valid for 12V power
Scenario
Instruments
1
Analog
2
Analog
3
Analog
4
Analog
5
Smart digital 1x
6
Smart digital 2x
7
Smart digital 4x
8
Smart digital 4x
9
Smart digital 4x
Tabel updated 20-03-2018 Valid for power from 6x AA battery
Scenario
Instruments
1
Analog
2
Analog
3
Analog
4
Analog
5
Smart digital 1x
6
Smart digital 2x
7
Smart digital 4x
8
Smart digital 4x
9
Smart digital 4x
Running Smart instruments on 6x AA battery is not practical.
Estimated days of working time = mAh of batteries *50% ÷ (mAh per day LogboxSE and instruments)
Measurement
GSM data
interval seconds
transfer
60
None
1
None
1
Every 24 hr
1
Every hr
1
None
1
None
1
None
1
Every 24 hr
1
Every hr
Measurement
GSM data
interval seconds
transfer
60
None
1
None
1
Every 24 hr
1
Every hr
1
None
1
None
1
None
1
Every 24 hr
1
Every hr
8
LogboxSE and instrument
current in mA
0.15
1.5
1.51
3.2
7
11
19
19.07
20.7
LogboxSE and instrument
current in mA
0.15
1.5
1.51
3.2
11
16
35
35.07
36.7
mAh per day LogboxSE
and instruments
3.6
36
37.7
76
168
264
456
457
496
mAh per day LogboxSE
and instruments
3.6
36
37.7
76
264
456
840
841
880

Advertisement