Setup Flow - Kipp & Zonen LOGBOX SE Instruction Manual

Table of Contents

Advertisement

ModemMode selects between sending the log files as email attachments or to an FTP server.
All other settings are for future use.
SMS Center is the number of your GSM provider and in international format. E.g. +31 6 5432 9876
When the modem is programmed to be on and the SMS Center is set and an SMS with 'Data?' is
received, the last measured data is send back as an SMS.
The following settings are standard settings when you don't have your own SMTP email gateway.
APN internet is unique for your GSM provider but very often internet is used
IP smtp.webglobe.sk
EmailFrom sender@physicus.eu
EmailTo your.email@your-company.com
UserName sender@physicus.eu
Password Sender39
TimeServer sk.pool.ntp.org
If you do have your own smtp or ftp provider you have to change IP, UserName and Password to your login.
The LOGBOX SE with 2G only does not allow for secure connection or secure authentication. The newer model does
support it.
If your email provider doesn't even need user authentication for SMTP then leave UserName and Password empty.
If you want to send the log files to an FTP server than the IP, UserName and Password need to reflect your FTP server login.
User settings on your FTP server need to be set correct and allow for writes for by the LOGBOX SE.

9.9 Setup flow

When starting with a LOGBOX SE it is best to use the following sequence:
1.
Write your equipment and attachments to the LOGBOX SE using e.g. Planned Setup on page 45.
2.
Make the analog instruments known using the first tab called Analog. Don't forget to tick the Measure box.
3.
Make the digital signals (On, Off, Frequency) known using the Digital tab
4.
Make the RS-485 Modbus instruments known using the Serial tab
5.
Select which input to log and in which order using the Output tab
Select when needed the Min, Max and StdDev to be logged as well
Select the right polynomials. At least A1 =1 is needed to log the value unprocessed.
6.
Select the measurement and logging interval using the System tab
DataFormat (Space delimited or Comma delimited CSV) can be selected
7.
If you do have a SIM card and data capabilities the systems can be personalised using the Modem tab
8.
To check if the logging is working the LiveView tab can be used.
For just checking it is advised to use short measurement and logging intervals and finally change them to the desired
values when ready.
28