OHAUS MB120 Service Manual page 79

Moisture analyzer
Hide thumbs Also See for MB120:
Table of Contents

Advertisement

Fill in the below and click Next.
1. Analyzer Serial Number.
2. Analyzer Load Cell Serial Number.
3. Model Number.
4. Point to the Analyzer and Load Cell image file (eif files) provided by OHAUS.
5. Click Next to complete the Write Image File process.
Note: After completing the above information when you click 'Next' you will hear a beep sound
from the Analyzer and the Analyzer will display the below message. Wait until the download
process is completed then ONLY PRESS the 'OK' button on the Analyzer display in order for the
Analyzer to re-boot (a normal re-boot would take approximately 20-30 sec).
MB90 MB120 Series Service Manual
APPENDIX B SOFTWARE SERVICE TOOL INSTRUCTIONS
73
Ohaus Corporation www.ohaus.com