1.9.2.2 Temperature calibration
You must have a temperature calibration kit to perform this procedure. If the Moisture Analyzer
has been recently used, allow at least 30 minutes before performing calibration.
Note: The temperature calibration kit is available as optional equipment.
Note: The unit will not calibrate with the pan support in place.
MB90 MB120 Series Service Manual
CHAPTER 1 - GETTING STARTED
Press Setup button on the main
screen.
Press Weight and temperature
calibration button.
Press Temperature calibration
button.
You are now prompted to remove
the pan handler and pan support.
Replace the pan handler and place
a temperature calibration unit on
the pan handler.