CHAPTER 3 – MAINTENANCE PROCEDURES
3.11.1.4 Replacing the Thermal Overload.
1. Open the lid (3.11.1).
2. Remove the Temperature sensor (3.11.1.1)
3. Remove the halogen lamp if necessary (3.11.1.3)
4. Remove the two wires connected to the Thermal overload.
5. Use a wrench to loosen the nut which securing the Thermal overload in place.
MB90 MB120 Series Service Manual
47
Ohaus Corporation www.ohaus.com