Test leck for sniffer leak testing

Advertisement

Quick Links

O P E R A T I N G I N S T R U C T I O N
lina80e1-06 (1603)
Test Leck for Sniffer
Leak Testing
Catalog No.
122 20 bis 122 23S
122 25, 122 25S
122 27 bis 122 43
122 75
123 22

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 122 20 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Inficon 122 20

  • Page 1 O P E R A T I N G I N S T R U C T I O N lina80e1-06 (1603) Catalog No. 122 20 bis 122 23S 122 25, 122 25S 122 27 bis 122 43 122 75...
  • Page 2: Table Of Contents

    Content Description General 1.1.1 Purpose 1.1.2 Validity Safety information Technical data Operation Description Handling Maintenance Cleaning the outside Repair Returning the Product Decommissioning Content...
  • Page 3: Description

    Description General INFICON covers no responsibility nor warranty if the user or third parties • disregard the information in this document • utilize the product not according to the defined use • make any kind of changes (modifications, alterations, etc.) to the product.
  • Page 4: Purpose

    1.1.2 Validity This manual applies to products with Cat. No. Catalog number Gas of test leak 122 20 TL R134a 122 20S TL R134a 122 20SN TL for special gases (non flammable) 122 20SF...
  • Page 5: Safety Information

    Catalog number Gas of test leak 122 34 Halon 1301 122 35 TL HFO-1234yf 122 36 TL R32 122 37 S-TL 4 Helium 122 37S S-TL 4 Helium 122 38 S-TL 5 Helium 122 39 S-TL 6 Helium 122 40 TL R134a 122 40SN TL for special gases (non flammable)
  • Page 6: Technical Data

    Calibration gas: Avoid eye-contact and inhalation in high con- centrations. Pass on the safety information to other users. Technical data Flammable: Keep away from ignition sources. Smallest deviation from calibration ± 10 % pressure range 0.5 bar to 8 bar temperature range +15°C to +35°C Storage temperature...
  • Page 7 0,07 … 0,18 122 26 R 32 2… 8 0,07 …0,28 122 36S R 134a 2 … 5 0,07 …0,18 122 20 R 134a 2 … 11 0,07 ... 0,39 122 20S R 134a 10 … 14 0,36 … 0,49 122 40 R 152a 2 …...
  • Page 8 Gas type Leak rate, typical Cat. No. oz/y HFO- 2 … 5 0,07…0,18 122 35 1234yf R 1301 2 … 17 0,07 … 0,60 122 34S R744 2 … 3,5 0,07…0,12 122 32 R744 10 … 14 0,36…0,50 122 75 1,00 - 1,20 x 10 mbar l/s 122 37...
  • Page 9: Operation

    Operation Description The calibrated test leak is installed in a rectangular housing. The front view is shown in Fig. 2. Pos. Description Manometer Leak rate of the calibrated test leak. (Example: Gas R152a, leak rate 2,64 g/a) Calibration gas outlet Date of calibration Fig.
  • Page 10: Maintenance

    When the calibration gas is used up (pressure < 0.5 bar) the test leak has to be refilled and calibrated. Please contact your nearest INFICON Service Center. Notice: If the pressure is < 0.5 bar a proper calibration of the leak detector is not assured.
  • Page 11: Repair

    Repair Defective products must not be opened. Please contact your nearest INFICON-Service Center. Returning the Product Calibration gas Test leaks sent for repair may still contain calibration gas. Take account of the national regulations for shipping the respective calibra- tion gas.
  • Page 12 INFICON GmbH, Bonner Strasse 498, D-50968 Cologne, Germany UNITED STATES TAIWAN JAPAN KOREA SINGAPORE GERMANY FRANCE UNITED KINGDOM HONG KONG V is i t ou r w eb s it e f or c on t ac t i n f o r m at i o n a nd s a l e s o f f i c e s .

Table of Contents