Inficon TC1000 Operating Manual
Inficon TC1000 Operating Manual

Inficon TC1000 Operating Manual

Test chamber for hermetically sealed parts

Advertisement

Quick Links

O P E R A T I N G
iina11e1-a (0604)
TC1000
test chamber for hermetically sealed
parts
M A N U A L
Catalog No.
551-005

Advertisement

Table of Contents
loading

Summary of Contents for Inficon TC1000

  • Page 1 O P E R A T I N G M A N U A L iina11e1-a (0604) Catalog No. 551-005 TC1000 test chamber for hermetically sealed parts...
  • Page 2: Table Of Contents

    50958 Köln Germany About these instructions These instructions are intended for users of the TC1000. The aim of the instructions is to provide you with a rapid introduction to assembly and operation of the test chamber. The structure of the instructions enables you to read the instructions from start to finish, or only the sections relevant to you if you are already familiar with the unit.
  • Page 3 3.2.1 Setting the measuring time 3.2.2 Setting the verification limit (Trigger level 1) 3.2.3 Serial errors when testing 3.2.4 Counting tested components 3.2.5 Reducing the helium substrate (reference measurement) Maintenance schedule Displaying and reading out the warning table Further documents Replacement parts list Decontamination declaration...
  • Page 5: Test Chamber Tc1000

    Test chamber TC1000 The TC1000 is a small vacuum chamber which allows you to carry out leak tests on sealed, small solid bodies. These include e.g. quartzes, relays, sensors, ICs and LEDs. You can easily mount the test chamber on the leak detectors UL1000 and UL1000 Fab.
  • Page 6: Technical Data

    Warning triangle Indicates a upcoming warning message which you can see by pressing the button. Technical data Diameter 130 mm Depth 40 mm Volume approx. 430 cm³ Weight 2.7 kg Connecting flange DN 25 KF Test chamber TC1000...
  • Page 7: Structure Of The Tc1000

    Structure of the TC1000 The test chamber Abb. 2-1 The closed test chamber Item Description Cover The cover of the test chamber has a floating mounting. It rests securely on the sealing lock. Connecting cable The electrical connection to the leak detector.
  • Page 8: The Test Chamber On The Leak Detector

    The bore holes are already provided for this. The test chamber on the leak detector The test chamber can preferably be operated on the INFICON leak detectors UL1000 or UL1000 Fab (software version 4.0 or higher). In the following we describe use of the test chamber on a UL1000 Fab.
  • Page 9: The Test Leak Adapter

    Abb. 2-4 If you carry out an external calibration, you will need an adapter cable and a test leak, e.g. leak test TL7. Section 3.1.2.4, Calibrating the leak detector describes how to calibrate the leak detector. Structure of the TC1000...
  • Page 10: Scope Of Supply

    Scope of supply The delivery comprises: • the test chamber TC1000, • an ESD bracelet, • the operating instructions for the unit (iina11d1) Accessories Catalogue number • ESD mat 551-002 • Test leak adapter 200 00 1797 • Test leak TL 7 (leakage rate in the range10 mbar l/s) 14210 •...
  • Page 11: Operating The Tc1000

    3.1.1 Unpacking Now remove the plastic casing in which the TC1000 and accessories are packed. Check the delivery for completeness. The scope of supply are listed in Section 2.4. Keep the plastic casing and packaging so that you can store the test chamber in it during periods when you are not using the test chamber.
  • Page 12: Setting The Auto Leak Test Operating Mode

    You can find further information on the Auto Leak Test operating mode in Section 3.1.2.2. 3.1.2.2 Setting the Auto Leak Test operating mode Switch on the leak detector. Press the Menu button. The Display main menu appears on the screen of the leak detector. Operating the TC1000...
  • Page 13: Setting The Test Chamber For The Test

    → V → A ETTINGS ACUUM ETTINGS EST SETTINGS The Auto Leak Test Settings menu now appears. To be able to work in measuring mode, you need to set at least the measuring time and the rejection rate. Operating the TC1000...
  • Page 14: Calibrating The Leak Detector

    Follow the instructions on the display. Calibrating with the external test leak: Connect a test leak with manual valve to the adapter cover. Open the cover of the test chamber for this and place the adapter cover on it. Operating the TC1000...
  • Page 15: Measuring Mode

    Press the STOP button. The leak detector interrupts the test mode. The leak detector does not show any result. 3.1.2.7 Interrupting and restarting the test mode Press the button next to the pictogram of the coffee cup. Operating the TC1000...
  • Page 16: Shutting Down

    Take out the cardboard box with the padding paper which was provided with the test chamber. Pack the TC1000 and accessories in the plastic casing. Place the TC1000 in the padded cardboard box. Settings in the Auto Leak Test menu...
  • Page 17: Setting The Measuring Time

    You can find further information at 3.1.2.4, Calibrating the leak detector. Examples of settings for time The INFICON test chamber was used with a content of 430 cm³ for the settings in these examples. Rejection rate Measuring time (recommended) 10E-5...
  • Page 18: Setting The Verification Limit (Trigger Level 1)

    → P AIN MENU ETTINGS ACUUM ETTINGS ETTINGS UMBER Enter the number of the first component. Afterwards the unit automatically counts up the number of tested components. The unit then counts the tested components and displays the result. Operating the TC1000...
  • Page 19: Reducing The Helium Substrate (Reference Measurement)

    Displaying and reading out the warning table A warning triangle appears on the display if faults occur. If you wish to read out the fault display, press the button next to it. Follow the instructions to rectify the warning. Operating the TC1000...
  • Page 20 3-10 Operating the TC1000...
  • Page 21: Further Documents

    This form can be downloaded Copies: from our website. Original for addressee - 1 copy for accompanying documents - 1 copy for file of sender INFICON GmbH Bonner Str. 498,50968 Cologne, Germany zisa01e1-a Tel: +49 221 3474 2222 Fax: +49 221 3474 2221 www.inficon.com leakdetection.service@inficon.com...
  • Page 22 INFICON GmbH, Bonner Strasse 498, D-50968 Cologne, Germany Phone: +49 (0)221 347-40 Fax: +49 (0)221 347-41429 E-mail: leakdetection@inficon.com UNITED STATES TAIWAN JAPAN KOREA SINGAPORE GERMANY FRANCE UNITED KINGDOM HONG KONG Vi s i t o ur w eb s i t e f o r c on ta c t i nf o rma t i on an d o t he r s al es of f i c e s w or l dwi de . w w w. i n fi c o n . com...

This manual is also suitable for:

551-005

Table of Contents