Using the IAQPoint
Programming Menus
Scroll to the CO
The next screen requests confirmation; Use the Up or Down
buttons to display YES and press Enter to confirm. (If there is no
confirmation within 32 seconds, the calibration is automatically
canceled and display returns to the main menu option.)
A confirmation screen is briefly displayed and the zero calibration
begins.
A success or failure message is displayed and the screen returns to the
main menu option.
Calibrating the CO
This option allows users to calibrate the sensor's span.
Note: When calibrating a duct mounted unit,it is best to remove the unit
from its base to ensure proper calibration.
Before starting the calibration (using the programming menus),
connect the regulator to the gas cylinder.
Adjust the flow rate to 0.1 L/min.(for the 2000 ppm range)
Connect the tubing from the regulator to the sensor, as shown in
the image at left.
26
Set Zero screen and press Enter to select.
2
CO
CO
CO
Sensor
2
Start exposing the sensor to the calibration gas at least
10 minutes before starting and throughout the calibration
process.
IAQPoint User Manual
Honeywell
* Menu *
Set Zero
2
Set Zero
2
YES
*Menu*
In Calib
Set Zero
2
SUCCESS