Honeywell IAQPoint User Manual page 29

Indoor air quality monitor
Hide thumbs Also See for IAQPoint:
Table of Contents

Advertisement

Available languages

Available languages

Let the gas flow for at least 10 minutes before starting the
calibration
Scroll to the "
select.
The next screen allows you to specify the span calibration gas
concentration value.
Use the Up or Down buttons to increase or decrease the value and
press Enter to select.
The next screen requests confirmation; press Enter to confirm and
the screen displays the following message. (If there is no
confirmation within 32 seconds, the calibration is automatically
canceled and display returns to the main menu option)
A success or failure message is displayed and the screen returns to the
main menu option.
511838
11/07
.
Set Span" main menu option and press Enter to
CO
2
IAQPoint User Manual
Honeywell
* Menu *
CO
Set Span
2
Set Span Gas
2000
CO
Set Span
2
YES
*Menu*
In Calib
CO
Set Span
2
SUCCESS
Using the IAQPoint
Programming Menus
27