Section 6

6.1 Troubleshoot the controller

6.2 Troubleshooting the analyzer

Message
REAGENT A LEVEL
REAGENT B LEVEL
STAND. SOLU. LEVEL
CLEAN. SOLU. LEVEL
Message
WARMUP PHASE
The analyzer is in warmup phase after
startup (de-frosting).
COOLING DOWN
The analyzer is decreasing temperature
after startup.
ANALYZER TOO
The analyzer temperature is less than 15°C
COLD
(59 °F).
ANALYZER TOO
The analyzer temperature is more than
WARM
55°C (131°F).
Troubleshooting
If entries are implemented with a delay or are not accepted for a short time, the delay
may be caused by a busy data network. Refer to the troubleshooting section in the
controller documentation.
After a software update, a system expansion or an interruption in the power supply, it may
be necessary to set the controller settings again. Record all the settings values that are
changed or entered so all the necessary data can be used to configure the parameters
again.
If, in usual operation, problems occur that are apparently caused by the controller, reboot
the controller as follows:
1. Save all important data that is on the SC Controller.
2. Remove power from the controller. Wait 5 seconds.
3. Supply power to the controller.
4. Make sure that all of the applicable settings are correct.
5. If the problems continue, contact technical support.
If the analyzer operation is not correct, identify if the humidity sensor has tripped. Repair
the damage, dry the humidity sensor and restart the system. If the problem continues,
contact technical support.
The analyzer has a status indicator light that shows different analyzer conditions, errors
and warnings. Refer to
Status indicator light
Reminders
Reminders show on the controller display. To see all of the reminders, push Menu then
select DIAGNOSTICS > PHOSPHAX sc LR > REMINDER LIST.
Possible cause
Level of reagent A is below the reminder level.
Level of reagent B is below the reminder level.
Level of standard solution is below the reminder
level.
Level of cleaning solution is below the reminder
level.
Warnings
Warnings show on the controller display. To see all of the warnings, push Menu then
select DIAGNOSTICS > PHOSPHAX sc LR > WARNING LIST.
Possible cause
on page 9.
Examine the reagent level and replace
the reagent if necessary. Refer to
Replace the reagents
set the reagent level counter to zero.
The system can only accurately
calculate the solution level if the counter
is reset each time the solution is
changed.
Solution
Wait until the warmup phase completes.
Wait until the cooling phase completes.
Identify if the heating system operation is correct.
Close the analyzer door.
Change/clean air filter. Identify if the fan operation is
correct.
Solution
on page 46. Then,
55