Installation
Location Filtration
Indoor
Continuous
sample feed

3.5.1 Plumbing one outdoor analyzer

18
Table 3 System configuration options (continued)
Drain
Number of
analyzers
Unheated 1
Unheated 2
This system configuration option uses one sc analyzer with the FILTRAX instrument for
the sample line. The waste from the analyzer is released into an open drain through the
optional heated drain hose.
Do the steps that follow to install one outdoor analyzer. Refer to
1. Install the FILTRAX into the sample stream. Refer to the FILTRAX User Manual for
more information.
2. Use a tubing plug to put the heated sample hose from the FILTRAX through the
analyzer access port.
3. Use a tubing plug to put the heated drain hose through the analyzer access port.
Note: The two samples lines of the heated drain hose are not used.
4. Use the sealing plug to close access ports that are not used.
5. Connect the heated drain connections. Refer to
on page 34.
6. Connect the heated drain tube to the sample outlet T-fitting.
Number of
Plumbing option
parameters
1
2
Refer to
Plumbing one
indoor analyzer with
continuous sample feed
on page 27.
Refer to
Plumbing two
indoor analyzers with
continuous sample feed
on page 27.
Figure
7.
Connect the optional heated drain