Plumbing One Outdoor Analyzer With Continuous Sample Feed - Hach PHOSPHAX sc LR User Manual

Table of Contents

Advertisement

Figure 8 Plumbing two outdoor analyzers
1 PHOSPHAX sc LR analyzer
2 Heated drain sample lines (not used)
3 AMTAX sc analyzer
4 Sample line to second analyzer (overflow vessel tube)
5 Drain tube cut from first analyzer

3.5.3 Plumbing one outdoor analyzer with continuous sample feed

This configuration option uses one outdoor sc analyzer and one sample preparation unit
that supplies a continuous sample stream. The waste from the analyzer is released into
an open drain through the optional heated drain hose.
Do the steps that follow to install one outdoor analyzer with continuous sample feed.
Refer to
Figure
9.
1. Install the sample preparation unit.
2. Use a tubing plug to put the heated sample hose from the sample preparation unit
through the analyzer access port.
3. Connect the sample line to the bottom inlet on the overflow vessel.
6 T-fitting from first analyzer
7 Heated drain
8 Heated drain from first analyzer
9 FILTRAX sample line
Installation
21